Appendix C: Communications via PROFIBUS
C.10 Diagnostics
SITRANS LR250 (PROFIBUS PA)
242 Operating Instructions, 01/2014, A5E32221386-AB
Acyclic Extended Diagnostics /General Fault Codes
LCD display
Meaning
Corrective Action
Byte
Bit
S:12
* Internal temperature of the device has
exceeded
specifications: it is operating outside its
temperature range.
Relocate device and/or lower process
temperature enough to cool device.
Inspect for heat-related damage and
contact your local Siemens
representative if repair is
required.
Fault code will persist until a manual
reset is performed using PDM or the
LCD interface.
1
5
S:14
Upper and lower input values (Process
Value Scale) for AIFB1 are the same.
Check configuration for AIFB1.
Ensure that Upper Value and Lower
Value (Process Value Scale) are not
the same.
6
S:15
Upper and lower input values
(Process Value Scale) for AIFB2
are the same.
Check configuration for AIFB2.
Ensure that Upper Value and Lower
Value (Process Value Scale) are not
the same.
7
S:17
Calibration interval as defined in
Maintenance Required Limit has expired.
Perform calibration.
2
1
S:18
Calibration interval as defined in
Maintenance Demanded Limit has
expired.
Perform calibration. 2
S:28
Internal device failure caused by a RAM
memory error.
Repair required: contact your local
Siemens representative.
3
4
S:29
EEPROM damaged. Repair required: contact your local
Siemens representative.
5
S:31
Flash error. Repair required: contact your local
Siemens representative.
7