User's Manual
Table Of Contents
- 8 Parameter reference
- 9 Service and maintenance
- 10 Diagnosing and troubleshooting
- 11 Technical data
- 12 Dimension drawings
- 12.1 Threaded horn antenna
- 12.2 Threaded horn antenna with extension
- 12.3 Flanged horn antenna
- 12.4 Flanged horn antenna with extension
- 12.5 Flanged encapsulated antenna (2"/DN50/50A sizes only)
- 12.6 Flanged encapsulated antenna (3"/DN80/80A sizes and larger)
- 12.7 Threaded PVDF antenna
- 12.8 Threaded connection markings
- 12.9 Raised-Face flange per EN 1092-1 for flanged horn antenna
- 12.10 Raised-Face flange per EN 1092-1 for flanged encapsulated antenna
- 12.11 Flat-Face flange
- 12.12 Process connection tag (pressure rated versions)
- A Appendix A: Technical reference
- B Appendix B: PROFIBUS PA profile structure
- C Appendix C: Communications via PROFIBUS
- D Appendix D: Certificates and Support
- 13 List of abbreviations
- 14 LCD menu structure
- Glossary
- Index
Parameter reference
SITRANS LR250 (PROFIBUS PA)
152 Operating Instructions, 01/2014, A5E32221386-AB
Device Address (5.1.)
Note
The address can be changed and locked from a remote master. See
PROFIBUS address
(Page
231) for details on disabling the address lock and
Master Reset (4.1.)
to reset Device
Address to 126.
Sets the unique address of the device on the network (also called PROFIBUS address).
Values
0 - 126. Default: 126
To set Device Address via SIMATIC PDM:
● Open the project in
Process Device Network View
then right-click on the device.
● Go to
Object Properties > Connection
to access the field
Short Address.
To change Device Address via the handheld programmer:
See Device Address (Page 49)
for details.
PROFIBUS Ident Number (5.2.)
Identifies the device on the network. The Ident Number must match that in the GSD file (the
GSD file provides information on the device to the master).
Options
STD PROFILE
Standard Profile (uses generic GSD for 2 AIFB)
[ident # = 0x9701]
*
MANUFACTURER
Manufacturer-specific (uses Siemens EDD and
GSD file, which identifies the LR250 PROFIBUS
PA [ident # = 0x8150]
STD – AIFB 1
ONL.
Standard Profile AIFB 1 only (uses generic GSD
for 1 AIFB) [ident # = 0x9700]