User's Manual
Table Of Contents
- 8 Parameter reference
- 9 Service and maintenance
- 10 Diagnosing and troubleshooting
- 11 Technical data
- 12 Dimension drawings
- 12.1 Threaded horn antenna
- 12.2 Threaded horn antenna with extension
- 12.3 Flanged horn antenna
- 12.4 Flanged horn antenna with extension
- 12.5 Flanged encapsulated antenna (2"/DN50/50A sizes only)
- 12.6 Flanged encapsulated antenna (3"/DN80/80A sizes and larger)
- 12.7 Threaded PVDF antenna
- 12.8 Threaded connection markings
- 12.9 Raised-Face flange per EN 1092-1 for flanged horn antenna
- 12.10 Raised-Face flange per EN 1092-1 for flanged encapsulated antenna
- 12.11 Flat-Face flange
- 12.12 Process connection tag (pressure rated versions)
- A Appendix A: Technical reference
- B Appendix B: HART communications
- C Appendix C: Certificates and support
- 13 List of abbreviations
- 14 LCD menu structure
- Glossary
- Index
Diagnosing and troubleshooting
10.3 General fault codes
SITRANS LR250 (HART)
176 Operating Instructions, 01/2014, A5E32220602-AB
Code/
Icon
Meaning
Corrective Action
S: 36
* Unable to start microwave module. Cycle power. If fault persists, contact your local Siemens
representative.
S: 37
* Measurement hardware problem. Cycle power. If fault persists, contact your local Siemens
representative.
S: 38
* Failure in the device electronics. Cycle power. If fault persists, contact your local Siemens
representative: repair required.
S: 43
* Factory calibration for the radar receiver has
been lost.
Repair required: contact your local Siemens
representative.
S: 45
* No valid boot program detected: firmware
corrupt
Repair required: contact your local Siemens
representative.
S: 48
* User configuration is invalid. One or more of
parameters: Low Calibration Point, High
Calibration Point, Volume breakpoints, and/or
Auto False-Echo Suppression, are set to
invalid values.
• Reconfigure the unit.
• Ensure the difference between High Calibration Point
and Low Calibration Point is not less than or equal to
zero; do a Master Reset.
S: 49
* EEPROM corrupt. Repair required: contact your local Siemens
representative.
S: 50
* EEPROM corrupt. Repair required: contact your local Siemens
representative.
S: 51
* EEPROM corrupt. Repair required: contact your local Siemens
representative.