10.3 General fault codes
SITRANS LR250 (HART)
Operating Instructions, 01/2014, A5E32220602-AB
177
Code/
Icon
Meaning
Corrective Action
S: 52
Fail-safe is activated. Possible causes:
1. hardware failure
2. memory failure
3. Fail-safe LOE timer expiredpossible
causes: faulty installation, antenna material
buildup, foaming/other adverse process
conditions, invalid calibration range.
For 3:
Correct configuration; ensure installation is correct;
no antenna buildup;
adjust process conditions to minimize foaming/other
adverse conditions;
correct calibration range.
If fault persists, or for 1 and 2, contact your local Siemens
representative.
S: 53
* Configuration lost: one or more parameter
settings have been lost. This may occur after a
firmware upgrade causes user parameters to
be reset.
Restore user parameters using SIMATIC PDM.