Parameters
To set Auto False Echo Suppression via SIMATIC PDM:
Open the menu Device – Echo Profile Utilities and click on the tab Auto
False Echo Suppression.
(For more detailed instructions see
Auto False Echo Suppression
on page
52.)
To set Auto False Echo Suppression via the handheld programmer:
c) Go to Auto False Echo Suppression Range (2.8.7.2.) and enter the
value calculated in step b).
d) Go to Auto False Echo Suppression (2.8.7.1.) and press RIGHT arrow
to open Edit Mode
e) Select Learn. The device will automatically revert to On (Use
Learned TVT) after a few seconds.
Options
OFF Default TVT will be used.
* ON ’Learned’ TVT will be used.
LEARN ’Learn’ the TVT.
After Auto False Echo Suppression
learned TVT
material level
false echo
Auto False Echo Suppression
Range
echo
marker
Level (db)
default