User's Manual
Table Of Contents
- Table of Contents
- SITRANS LR250 Overview
- Specifications
- Installation
- Wiring
- Operating via the handheld programmer
- Operating via SIMATIC PDM
- Functions in SIMATIC PDM
- Quick Start Wizard via SIMATIC PDM
- Changing parameter settings using SIMATIC PDM
- Parameters accessed via pull-down menus
- Operating via AMS Device Manager
- Functions in AMS Device Manager
- Features of AMS Device Manager
- Device Description (DD)
- Configuring a new device
- Startup
- Pull-down menu access
- Device configuration
- Quick Start Wizard via AMS Device Manager
- Maintenance and Diagnostics
- Remaining Device Lifetime [see Remaining Device Lifetime (4.2.) on page 99]
- Remaining Sensor Lifetime [see Remaining Sensor Lifetime (4.3.) on page 102]
- Service Schedule [see Service Schedule (4.4.) on page 105]
- Calibration Schedule [see Calibration Schedule (4.5.) on page 108]
- Electronic Temperature
- Wear (see Wear on page 56)
- Communication
- Security
- Device Diagnostics
- AMS Menu Structure
- Functions in AMS Device Manager
- Parameter Reference
- 1. Quick Start
- 2. Setup
- 3. Diagnostics
- 4. Service
- 5. Communication
- 6. Security
- 7. Language
- Appendix A: Alphabetical Parameter List
- Appendix B: Troubleshooting
- Appendix C: Maintenance
- Appendix D: Technical Reference
- Principles of Operation
- Echo Processing
- Analog Output
- Maximum Process Temperature Chart
- Process Pressure/Temperature derating curves
- Loop power
- Appendix E: Application Examples
- Appendix F: HART Communications
- Appendix G: ATEX Certificates
- Appendix H: Firmware Revision History
- Glossary
- Index
- LCD menu structure
7ML19985JE03 SITRANS LR250 (HART) – INSTRUCTION MANUAL Page 7
mmmmm
Specifications
Specifications
Power
• Maximum 30 V DC
• 4 to 20 mA
• Max. startup current see
Startup Behavior
on page 141
Performance
Reference operating conditions according to IEC 60770-1
• ambient temperature +15 to +25 °C (+59 to +77 °F)
• humidity 45% to 75% relative humidity
• ambient pressure 860 to 1060 mbar g (86000 to 106000 N/m
2
g)
Measurement Accuracy (measured in accordance with IEC 60770-1)
• Maximum measured error = 5 mm (0.2") (including hysteresis and non-
repeatability)
Frequency K-band
Max. measurement range
1)
20 m (65.6 ft) except for 1.5" horn which has 10 m
(32.8 ft)
Min. detectable distance 50 mm (1.97") from end of horn
2)
Update time
3)
minimum 1 second, depending on settings for
Response Rate (2.4.1.) and LCD Fast Mode (4.9.)
Influence of ambient temperature < 0.003%/K (average over full temperature range,
referenced to maximum range)
Note: Siemens Milltronics makes every attempt to ensure the accuracy of these
specifications but reserves the right to change them at any time.
General Purpose
Non-incendive (FM/CSA US/Canada only)
Intrinsically Safe
Nominal 24 V DC at 550 Ohm
Flameproof
Increased Safety
Explosion-proof (FM/CSA US/Canada only)
Nominal 24 V DC at 250 Ohm
1)
From sensor reference point: see
Dimensions
on page 12 and
Flanged Horn with
extension
on page 14.
2)
Minimum range is horn length + 50 mm. See
Dimensions
on page 12.
3)
Reference conditions: Response Rate (2.4.1.) set to FAST, LCD Fast Mode (4.9.)
set to ON.