Page 140 SITRANS LR250 (HART) – INSTRUCTION MANUAL 7ML19985JE03
mmmmm
D: Technical Reference
Loop power
Typical Connection Drawing
Allowable operating area of SITRANS LR250
Curve 1 (General Purpose, Intrinsically Safe, Non-incendive)
Loop Voltage versus Loop Resistance
Note: Loop voltage is the voltage at the terminals of the power supply (not the voltage
at the terminals of the device).
+
-
LR250
loop resistance
RL
loop current IL
loop voltage
VL
power
supply
Power Supply Requirements
0
100
200
300
400
500
600
700
800
900
15 17 19 21 23 25 27 29
Loop Voltage - V
L
(Source Voltage)
Loop Resistance - R
L
(30.0V, 845 ohms)
RL = 44.6 * VL - 493.4
(17.8V, 300 ohms)
(16.7V, 0 ohms)
RL = 272.7 * VL - 4554.1
Loop Voltage – V
L
Loop Resistance – R
L
ALLOWABLE
OPERATING AREA
Power Supply Requirements