User's Manual

Table Of Contents
Page 26 SITRANS LR250 (HART) – INSTRUCTION MANUAL 7ML19985JE03
mmmmm
Wiring
1. Intrinsically Safe wiring
For power demands see
Curve 1 (General Purpose, Intrinsically Safe, Non-
incendive)
on page 140.
For wiring requirements: follow local regulations.
Approved dust-tight and water-tight conduit seals are required for outdoor
NEMA 4X / type 4X / NEMA 6, IP67, IP68 locations.
Refer to
Instructions specific to hazardous area installations
on page 30.
Recommended intrinsically safe barriers are listed under
Passive Shunt Diode
Barriers
on page 27 and
Active barriers (repeating barriers)
on page 27.
Note: Selecting a suitable PLC input module, power supply, or barrier requires
knowledge about Intrinsic Safety and the application. It is the responsibility of the
installer to ensure that the intrinsically safe installation complies with both the
apparatus approval requirements and the relevant national code of practice.
WARNING: Possible Static Hazard, Do Not Rub Or Clean On Site.
Ii = 120 mA
Ui = 30V
Li = 0.1mH
Pi = 0.8W
Ci = 15nF
HART
II1DIP68Txx
II 1 G EEx ia IIC T4
03ATEX2142X
IECEx SIR 05.0031X
Ex ia IIC T4
SIRA
SITRANS LR 250
Made in Canada
Siemens Milltronics Process Instruments Inc., Peterborough
7ML1234-56789-0ABC-D
Encl.: NEMA / TYPE 4X, 6, IP67, IP68
Amb.Temp.: 40 °C to 80 °C
Power Rating: 24V
Max., 4 - 20mA
Nom., 30 V
Serial No: GYZ / S1034567
The ATEX certificate listed on the nameplate can
be downloaded from the product page of our website at: www.siemens.com/LR250
.
See also
ATEX Certificate SIRA 06ATEX2358X
on page 150
The IECEx certificate listed on the nameplate can be viewed on the IECEx website. Go to:
http://iecex.iec.ch and click on Ex Equipment Certificates of Conformity then enter the
certificate number IECEx SIR 05.0031X.
WARNING: Possible Static Hazard, Do Not Rub Or Clean On Site.
Exia per drawing: 23650653
Class I, Div1,Group A,B,C,D
Class III
Class II, Div1,Group G
CANADA: 267P - LR 250
HART
R
FCCID:NJA-LR250
P
max
= 0.8 W
C
i
=15nF
Temp.Code:T4
V
max
= 30V
L
i
=0.1mH
I
max
= 120mA
SITRANS LR 250
Made in Canada
Siemens Milltronics Process Instruments Inc., Peterborough
7ML1234-56789-0ABC-D
Encl.: NEMA / TYPE 4X, 6, IP67, IP68
Amb.Temp.: 40 °C to 80 °C
Power Rating: 24V
Max., 4 - 20mA
Nom., 30 V
Serial No: GYZ / S1034567
The FM/CSA Intrinsically Safe connection drawing
number 23650653 can be downloaded from
the product page of our website at: www.siemens.com/LR250
.