User's Manual
Table Of Contents
- Table of Contents
- SITRANS LR250 Overview
- Specifications
- Installation
- Wiring
- Operating via the handheld programmer
- Operating via SIMATIC PDM
- Functions in SIMATIC PDM
- Quick Start Wizard via SIMATIC PDM
- Changing parameter settings using SIMATIC PDM
- Parameters accessed via pull-down menus
- Operating via AMS Device Manager
- Functions in AMS Device Manager
- Features of AMS Device Manager
- Device Description (DD)
- Configuring a new device
- Startup
- Pull-down menu access
- Device configuration
- Quick Start Wizard via AMS Device Manager
- Maintenance and Diagnostics
- Remaining Device Lifetime [see Remaining Device Lifetime (4.2.) on page 99]
- Remaining Sensor Lifetime [see Remaining Sensor Lifetime (4.3.) on page 102]
- Service Schedule [see Service Schedule (4.4.) on page 105]
- Calibration Schedule [see Calibration Schedule (4.5.) on page 108]
- Electronic Temperature
- Wear (see Wear on page 56)
- Communication
- Security
- Device Diagnostics
- AMS Menu Structure
- Functions in AMS Device Manager
- Parameter Reference
- 1. Quick Start
- 2. Setup
- 3. Diagnostics
- 4. Service
- 5. Communication
- 6. Security
- 7. Language
- Appendix A: Alphabetical Parameter List
- Appendix B: Troubleshooting
- Appendix C: Maintenance
- Appendix D: Technical Reference
- Principles of Operation
- Echo Processing
- Analog Output
- Maximum Process Temperature Chart
- Process Pressure/Temperature derating curves
- Loop power
- Appendix E: Application Examples
- Appendix F: HART Communications
- Appendix G: ATEX Certificates
- Appendix H: Firmware Revision History
- Glossary
- Index
- LCD menu structure
Page 84 SITRANS LR250 (HART) – INSTRUCTION MANUAL 7ML19985JE03
mmmmm
Parameters
2.6. Analog Output Scaling
2.6.1. Current Output Function
Can be set to either Level, Space, Distance, or Volume. (The device can carry
out a volume calculation only after a vessel shape has been specified.)
To view the mA reading in the secondary region of the LCD, press
on the
handheld programmer.
To modify Current Output Function via SIMATIC PDM:
• Open the menu Device – Select Analog Output. (See
Select Analog Output
on page 55 for more detail.)
Notes:
• Level, space, and distance, have different reference points.
• Use caution when changing Current Output Function while the device is
connected to a HART network. Current Output Function controls the
primary value and the loop current for the device.
• Current Output Function also affects the secondary, tertiary and
quaternary variables in a HART network.
Options Reference point Description
* Level Low Calibration Point measured as a percentage of the
difference between High Calibration Point
and Low Calibration Point
Space High Calibration Point
Distance sensor reference point
measured as a percentage of Low
Calibration Point
Volume converts a level measurement to a volume output
Manual
a)
(LUI only)
a)
Current Output Function must be set to Manual before modifying mA Output Value
(2.6.6.). Remember to restore the previous Current Output Function setting after
modifying mA Output Value.
allows mA Output Value (2.6.6.) to be set to a user-defined value
Level Space Distance
Distance
Space
Level
20 mA
100%
4 mA
0%
High
Cal.
Point
4 mA
0%
20 mA
100%
4 mA
0%
20 mA
100%
Sensor
ref.
pt.
Low
Cal.
Pt.