Start Up
Start Up
Self-test
The device performs a self-test after switching on the power supply. It is ready for
operation and programming when the LCD displays the multi-display and the control
elements can be operated.
Multi-display
The multi-display shows on the LCD after a successful self-test with the output of the
level in the first line and the signal-to-noise ratio in the second line (factory setting):
Local Programming
When the mult-display shows on the LCD, you are ready to begin local programming
using the optical control elements. To access the parameter settings, touch the
element once. "Main Menu" is displayed in the first line of the LCD. Then program the unit
beginning with the Auto-Setup parameters.
Auto-Setup
After switching on the SITRANS LR 400, and after a successful self test, set the Auto-
Setup parameters to make the system operational.: (see page 25).
The language of the local user interface (English or Deutsch)
The unit of length of the measured level
The nozzle height in the selected unit of length
•The vessel height in the selected unit of length)
The LRV as a distance from the bottom of the vessel
The URV as a distance from the bottom of the vessel
The damping of the measured level in s.
The application type
The bus-address by Profibus PA communication (on Profibus models)
Enter the necessary values as described in "Parmeters" on page 25.
If the multi-display does not appear or displays incorrect measured values after Auto-
Setup, proceed as described in “Troubleshooting” on page 62.
Refer to the Parameter section that begins on page 25 for a list of available parameters.
Note: Frequent switching off and on of the devices causes aging of the electronics
(see Parameter 3.1).