7ML19985JM01 SITRANS LR 460 – INSTRUCTION MANUAL Page 13
mmmmm
Installation
Installation
Mounting Location
Recommendations
Provide easy access for viewing display and programming via handheld
programmer.
Provide an environment suitable to the housing rating and materials of construction.
Mount the unit more than 1 meter away from vessel walls, pipes and other
assemblies as well as the filling stream, which could cause reflective interference.
Align the antenna so that the radar cone intersects the surface of the measuring
medium vertically if possible.
WARNINGS:
SITRANS LR 460 is to be used only in the manner outlined in this
manual, otherwise protection provided by the equipment may be
impaired.
Installation shall only be performed by qualified personnel and in
accordance with local governing regulations.
Improper installation may result in loss of process pressure
Notes:
Installation must be according to ETSI EN 302372.
Refer to device nameplate for approval information.
For Type 4X/NEMA 4X, Type 6/NEMA 6, IP67 type of protection use only approved,
suitable sized hubs for watertight applications.
Observe all maximum permissible ambient and process temperatures. Refer to
Maximum Process Temperature Chart
on page 86.
Provide a warning sign and/or touch guard if the surface of the measuring
instrument can become hotter than 70 °C (158 °F) in use.