Installation
Easy Aimer Installation
Easy Aimer Positioning
Direct SITRANS LR 460 so the horn antenna
is pointed at an angle to the filling level as
shown.
Loosen the Easy Aimer ball locking bolts and
position the LR 460.
Tighten bolts to secure the device.
top clamping plate (upper socket)
bottom clamping plate (lower socket
)
customer gasket as required
[recommended thickness 1.5 to
1.8 mm (0.06 to 0.07")]
customer mounting plate, as
required
ø 4" (102 mm) min., central
opening
30°
max.
2" horn: 6.5" (165.5 mm)
3" horn: 9.0" (228.3 mm)
4" horn: 11.5" (291.2 mm)
0.38"
(10 mm)
Easy Aimer ball locking bolts
S
IT
R
A
N
S
L
R
4
0
0
Easy Aimer