User's Manual

Table Of Contents
Page 94 SITRANS LR560 (FF) – INSTRUCTION MANUAL 7ML19985LY01
mmmmm
Parameters
2.4. Signal Processing
In AMS Device Manager, see the General tab under Signal Processing (LTB)
on page 53.
2.4.1. Near Range
The range in front of the device (measured from the sensor reference point
1)
)
within which any echoes will be ignored. (This is sometimes referred to as
"Blanking" or "Dead Zone".)
2.4.2. Far Range
Allows the material level to drop below Low Calibration Point without
generating a Loss of Echo (LOE) state. See CLEF Range (2.4.5.4.) for an
illustration.
Use this feature if the measured surface can drop below the Low Cal. Point in
normal operation. For more detail see
Far Range (2.4.2.)
on page 143.
1)
See
SITRANS LR560 with stainless steel universal flat flange
on page 12.
Values
Range:
Min = 0
Max. = 45 m (40 m device)
= 105 m (100 m device)
Default: 0.278 m (0.91 ft)
Related
parameters
Unit (2.3.1.)
Notes:
Far Range can extend beyond the bottom of the vessel.
When Low Calibration is updated, Far Range will be updated to Low
Calibration Pt. + 5 m (16.40 ft).
Values
Range:
Min. = 0
Max. = 45 m (40 m device)
= 105 m (100 m device)
Default = Low Calibration Pt. + 5 m (16.40 ft)
Related
Parameters
Unit (2.3.1.)
CLEF Range (2.4.5.4.)