User's Manual

Table Of Contents
Page 114 SITRANS LR560 (FF) – INSTRUCTION MANUAL 7ML19985LY01
mmmmm
Parameters
4.3.3. Remaining Lifetime
Read only. Lifetime Expected (4.3.1.) less Time in Operation (4.3.2.).
4.3.4. Reminder Activation
Allows you to enable a maintenance reminder.
a) First set the limit values in Reminder 1 (Required) (4.3.5.)/Reminder 2
(Demanded) (4.3.6.).
b) Select the desired Reminder Activation option.
4.3.5. Reminder 1 (Required)
If Remaining Lifetime (4.3.3.) is equal to or less than this value, the device
generates a Maintenance Required reminder.
a) Modify limit values as required.
b) Set Reminder Activation (4.3.4.) to the desired option.
4.3.6. Reminder 2 (Demanded)
If Remaining Lifetime (4.3.3.) is equal to or less than this value, the device
generates a Maintenance Demanded reminder.
a) Modify limit values as required.
b) Set Reminder Activation (4.3.4.) to the desired option.
4.3.7. Maintenance Status
Indicates which level of maintenance reminder is active.
To display the level of maintenance reminder in AMS Device Manager see
Extended Diagnostics (LTB) on page 70 under
Device Diagnostics (Level
Transducer Block - LTB)
.
4.3.8. Acknowledge Status
Indicates which level of maintenance reminder has been acknowledged.
Options
REMinder 1 (Maintenance REQuired)
REMinder 2 (Maintenance DEManded)
REMinders 1 AND 2 (Maintenance Required and
Maintenance Demanded)
*OFF
Values
Range: 0 to 20 years
Default: 0.164 years
Values
Range: 0 to 20 years
Default: 0.019 years