User's Manual
Table Of Contents
- Safety Notes
- FCC Conformity
- CE Electromagnetic Compatibility (EMC) Conformity
- Industry Canada
- The Manual
- Technical Support
- SITRANS LR560 Overview
- Specifications
- Installation
- Wiring
- Local Operation
- Operating via AMS Device Manager
- Functions in AMS Device Manager
- Programming via AMS Device Manager
- Adding a new device
- Startup
- Configuring a new device
- Changing Block Modes
- Configure/Setup Parameters
- Transducer Block Parameters
- Level Transducer Block Parameters
- Operation (LTB)
- Simulation (Input)
- Setup (LTB)
- General
- Calibration
- Rate
- Signal Processing (LTB)
- Range
- Echo Select
- Sampling
- Echo Quality
- Filtering
- TVT Setup
- TVT Shaper 1
- TVT Shaper 2
- TVT Shaper 3
- Echo Profile Parameters (view only)
- Maintenance & Diagnostics (LTB)
- Sensor Lifetime
- Service Schedule
- Communication (LTB)
- Communication:
- Configure/Setup (Liquid Crystal Display Block-LCD)
- Configure/Setup (Diagnostic Transducer Block-DIAG)
- Configure/Setup (Resource Block - RESOURCE)
- Device Diagnostics (Level Transducer Block - LTB)
- Device Diagnostics (Level Control Device Block - LCD)
- Device Diagnostics (Diagnostic Transducer Block - DIAG)
- Device Diagnostics (Resource Block - RESOURCE)
- Password Protection
- AMS Menu Structure
- Parameter Reference
- 1. Quick Start
- 2. Setup
- 2.1. Identification
- 2.2. Device
- 2.3. Sensor
- 2.4. Signal Processing
- 2.4.1. Near Range
- 2.4.2. Far Range
- 2.4.3. Minimum Sensor Value
- 2.4.4. Maximum Sensor Value
- 2.4.5. Echo Select
- 2.4.6. Sampling
- 2.4.7. Echo Quality
- 2.4.8. TVT (Auto False Echo Suppression) Setup
- 2.4.9. TVT Shaper
- 2.4.9.1. Shaper 1-9
- 2.4.9.2. Shaper 10-18
- 2.4.9.3. Shaper 19-27
- 2.4.9.4. Shaper 28-36
- 2.4.9.5. Shaper 37-45
- 2.4.9.6. Shaper 46-54
- 2.4.9.7. Shaper 55-63
- 2.4.9.8. Shaper 64-72
- 2.4.9.9. Shaper 73-81
- 2.4.9.10. Shaper 82-90
- 2.4.9.11. Shaper 91-99
- 2.4.9.12. Shaper 100-108
- 2.4.9.13. Shaper 109-117
- 2.4.9.14. Shaper 118-120
- 2.5. AIFB 1
- 2.6. AIFB 2
- 2.7. Measured Values
- 2.8. Filtering
- 3. Diagnostics
- 4. Service
- 5. Communication
- 6. Security
- 7. Language
- Appendix A: Alphabetical Parameter List
- Appendix B: Troubleshooting
- Appendix C: Maintenance
- Appendix D: Technical Reference
- Principles of Operation
- Echo Processing
- Measurement Range
- Measurement Response
- Damping
- Loss of Echo (LOE)
- Temperature derating curves
- Appendix E: Communications
- Appendix F: Firmware Revision History
- Glossary
- Index
- LCD menu structure
Page 46 SITRANS LR560 (FF) – INSTRUCTION MANUAL 7ML19985LY01
mmmmm
AMS Device Manager
Configuring a new device
Configure a new device using the Quick Start Wizard. The Quick Start Wizard provides an
easy step-by-step procedure that configures the device for a simple application.
Please consult the operating instructions or online help for details on using AMS Device
Manager.
Quick Start Wizard via AMS Device Manager
Quick Start Wizard steps
Launch AMS Device Manager and double-click the device icon from Device Connection
View to open the startup screen.
1)
Navigate to Configure/Setup > Resource Block > Wizards > Quick Start.
• In the navigation window, click on the Quick Start steps in order.
• At each step, if you do not wish to change the default values in the dialog window
that opens, click on the icon for the next Quick Start step.
• If you modify a parameter in any step, the Apply button is activated. Click Apply to
write changes to the device.
Notes:
• The LR560 FF is shipped with RESOURCE and LTB blocks in Out of Service mode.
• If you complete the Quick Start Wizard via local operation the first time it is used, it
will automatically put the RESOURCE and LTB blocks into AUTO mode. (See
Quick
Start Wizard via the LDI push buttons
on page 36.)
• Except for the first time it is used, after completing the Quick Start wizard via local
operation, you must manually put the RESOURCE and LTB blocks into AUTO mode.
• After completing the Quick Start Wizard via AMS, you must manually put the
RESOURCE and LTB blocks into AUTO mode.
Notes:
• When performing a Quick Start via AMS, the Resource and LTB blocks must be in
Out of Service mode before any configuration changes
1)
can be written. (See
Changing Block Modes
on page 49.)
• After completing the Quick Start wizard via AMS, you must manually place the
RESOURCE block in Automatic mode. This will also change LTB to Automatic mode.
• Values set using the Quick Start Wizard via AMS are saved and recalled each time it
is initiated.
1)
Changes to parameters that affect the block output.