The operating modes can be selected by pressing the corresponding
button on the alarm box or by using the System Manager in the
syngo MR software. See: Software operator manual
Additionally a main circuit breaker is located in the control room
which should NOT be used during proper functioning of the MR
system. It switches off the overall system including cooling, which
leads to helium boil-off.
Starting the system (System On)
System start-up includes the following steps:
Switching on the MR system at the alarm box
Switching on the syngo MR Workplace
Checking the MR system components
Do not perform preliminary examination steps (e.g. moving the
patient table, connecting coils) at the MR system while starting
up the system.
Prior to starting the system, the patient table should be in the Home
Position.
After “system off” or Standby, wait at least 30 seconds before you
switch on the system again.