Please note that Siemens provides customers with the optical
trigger signal output for research purposes only. No devices
connected to this output have been tested by Siemens. Before
connecting devices to the MR suite using the optical trigger
signal output, they must be tested for safety by trained
personnel.
Before using devices in the proximity of the magnet, their non-
magnetic properties and clinical operation in the magnetic field
have to be confirmed.
The use of devices connected to the optical trigger signal output
has to comply with any applicable governmental or local hospital
Institutional Review Boards (IRBS).
Siemens will not be held responsible for the use of any device
and resulting consequences in connection with the optical trigger
signal output.
The system separator contains electronics components and cooling-
equipment to deliver an adequate cooling power for the system. If a
special cooling system (chiller) is installed, the system separator is not
necessary.
RF coils
The description, handling and quality measurements for RF coils are
included in a separate operator manual. See: Coil operator manual
syngo Acquisition Workplace
The workplace in the control room is known as syngo Acquisition
Workplace (syngo Acq WP). It includes the host processor with the
operating elements monitor, keyboard, and mouse.
An additional component of the syngo Acquisition Workplace is the
intercom. ( Page 74 Description)
System separator
3.1.4
3.2
MR system components 3
MAGNETOM Skyra
fit
| syngo MR E11
49
Print No. M6-04401.621.02.02.02