Tabletop is in Home position.
Brake for the castors is released.
Pull the emergency undocking handle to undock table.
The table can be removed from the system.
1 Press the emergency undocking handle back into its original
position.
2 Position the table with the docking nose in the docking station up
to the end stop. Then press the left pedal down fully with your
foot.
Laser light localizer
The laser light localizer facilitates correct patient positioning. The
laser light localizer is located on top at the entrance to the magnet
bore.
All laser-relevant locations at the MR system are identified by warning
labels affixed directly next to the laser opening.
For some examinations, a manual alignment of the laser light
localizer to the coil is not necessary. If all relevant data (e.g. patient
orientation, body height, examination) are registered, auto-
positioning is indicated on the Dot display at the magnet cover.
Using the laser light localizer
Anesthetized patients or patients who do not have an eye blink reflex
must be protected against the effects of the laser beam.
Resetting the emergency
undocking handle