User's Manual

Table Of Contents
A malfunction within the supervision system is detected.
A corresponding error message appears at the syngo Acquisition
Workplace.
Dialog box at the syngo Acquisition Workplace
(1) Scanner hardware malfunction
Functional problem within automatic shutdown mechanism.
Please restart software of MeasRecon. If the problem persists,
please call Siemens Service immediately.
Call Siemens Service.
The supervision sensitivity is potentially affected, but the
operation of your MR scanner is still possible.
Acting in case of a supervision
malfunction
3 MR system components
82
MR System | Operator Manual
Print No. M6-04401.621.02.02.02