User's Manual

Addendum
10 / 12 AXB4-310.621.01.02.02 Addendum
DRAFT
Accessory parts Please note that for some accessory parts, special instructions on cleaning are giv-
en in the corresponding chapters.
If no special reference is given, then the following applies generally:
Use a lukewarm detergent solution and a soft cloth for slight contamination.
Remove major contamination first with a cloth soaked in alcohol, then wipe off
with clear water.
Remove blood spots best of all with cold water.
Remove contrast medium spots best of all with warm water.
After using disinfectants, always wipe off with clear water.
Disinfection
To disinfect surfaces we recommend common surface disinfectants based on
aldehyde and/or ampholyteric surfactants.
Caution
Disinfectant sprays
Sprays can seep into the units.
(This can cause damage to electronic components and the formation of flamma-
ble mixtures of air/solution vapor.)
Use other disinfection methods.
Note The following active ingredient classes can be used:
Guanidine derivatives
Peroxide compounds
Organic acids
The following products should not be used:
–Virex TB
Terralin
All alcohol-based products
All phenol-based products
All chlorine-releasing products
The safety data sheets of the manufacturer provide detailed information on the