User's Manual

Addendum
8 / 12 AXB4-310.621.01.02.02 Addendum
DRAFT
Charging Station
The Wireless Remote Control is plugged in the Charging Station for charging the
batteries. It is recommended to charge the batteries every day after deployment.
Charging Station
If the Wireless Remote Control needs charging during a patient procedure, simply
plug it in the Charging Station.
Installation The Charging Station can be installed either at the wall or on the table.
Exchange of batteries
The batteries of the Wireless Remote Control must be exchanged once a year by
authorized and trained staff. Please call Siemens Service. Unauthorized exchange
may cause severe damage.
Note Never use another charging device than the Charging Station delivered with the
Wireless Remote Control.
Note For safety reasons, never touch the charging contacts and the patient simulta-
neoulsy when the charging station is connected to a socket (charging contacts are
permanently energized).