Abdomen
117
Hint
• You could repeat the same protocol by simply click-
ing the chronicle with the right mouse button for
“repeat“, e.g., when both non-contrast and contrast
studies are required.
AbdFast
kV 130
Effective mAs 60
Rotation time 0.8 sec
Slice collimation 5.0 mm
Slice width 8.0 mm
Feed/Rotation 20.0 mm
Pitch Factor 2.0
Increment 8.0 mm
Kernel B41s
CTDIVol
6.47 mGy
Effective dose Male: 2.59 mSv
Female: 3.29 mSv
Contrast medium IV injection
Start delay 25 – 30 sec.
Flow rate 2.5 ml/sec.
Total amount 80 ml
C2-025.630.01.01.02_APPLICATIONGUIDE_SPIRIT.book Page 117 Friday, April 8, 2005 9:55 AM