syngo Volume Evaluation
Prerequisites
The images must fulfill the following conditions:
Only images of one patient must be loaded for eval-
uation with Volume.
The scans must have been acquired with the same
table height and the same gantry tilt.
Only image data sets with the same slice orientation
must be used; (the selection should not include a
topogram image).
The loaded images must have the same zoom factor
and the same reconstruction center.
At least four images must be loaded before volume
evaluation can be performed.
The time range between the first and the last image
must not exceed one hour.
The data capacity of the volume must not exceed
1024 images with matrix dimensions of
512
x 512 pixels.