Pulse module
WZU-P2
Type
Voltage
Current
Classification per EN 1434-2
Voltage drop
Classification per EN 1434-2
Voltage drop
Dielectric strength
Open collector
Max. DC 30 V
Max. 30 mA
OB
Approx. 1.3 V at 20 mA
OC
Approx. 0.3 V at 0.1 mA
500 V
eff
against ground
M-bus module
WZU-MB G4
WZU-MI
M-bus slave interface
Voltage
Power consumption
Address
Baud rate
Polarity
Galvanic isolation
From meter
From pulse inputs
As per EN 1434-3 and EN 13757-2 /-3
Max. DC 50 V
1 M-bus load (1.5 mA)
Primary or secondary
300, 1200, 2400, 4800, 9600 baud
Any
Yes
No
M-bus module WZU-MI
Part of pulse input
Pulse inputs
Number of inputs
Module battery
Battery type
Battery voltage
Battery life
Pulse duration (Low)
Pulse duration (High)
Frequency
Polarity
Galvanic isolation
From meter
From M-bus inputs
Pulse value
Display
Output voltage
Internal resistance
As per EN 1434-2, class IB
2
1 BR2477A (LiCF)
3 V
Operating: Up to 5 years
Storage time: +1 year
If M-bus voltage is fed to the module for at
least 90% of the operating time, battery life
is extended to 10 years
≥50 ms
≥50 ms
Max. 10 Hz
Yes
Yes
No
0.01 to 10,000.00 liters per pulse
In increments of 0.01 liter per pulse
In m3, 7 digits, without or with 1 decimal
place, depending on parameter setting
Approx. 3.3 V
Approx. 1 MΩ
Approx. 3 μA