Data Sheet for Product

Page 2 of 10
Siemens Industry, Inc
Specifications
Materials (within air stream) Standard
Duct Casing
mechanically locked and
sealed seam
Air Flow Sensor
galvanized steel back support
Liner
free foam which meets
requirements of NFPA 90A and
UL181. Closed cell structure
foam allows for disinfecting and
hand washing with detergents
and water.
Damper Shafts (2)
1/2-inch (1.27 cm) diameter,
zinc-plated steel. Shaft end is
marked with the damper blade
position.
Damper Blades (3)
seal sandwiched between two
22 gauge galvanized steel
blades mechanically locked
together
Damper Bushings
Materials (outside air stream) Standard
Control Enclosure
NEMA-1 18 gauge galvanized
steel
Pneumatic Tubing
UL rated 94 V-2 fire retardant
polyethylene
Pneumatic Fittings
Flow Measurement
Sensor Type
Four quadrant 12 multi-point,
center averaging and signal
amplification
Accuracy (terminal only)
Flow Accuracy
±2% of actual flow @ listed
ranges (sensor only, does not
include accuracy of controller
or transmitter)
Airflow Control and Temperature Control
Damper Blades
Round sealing single blade
with 90-degree control and 2
mechanical stops
Environmental
Operating
Temperature
% RH
40°F to 120ºF (0 to 50ºC)
0% to 95% non-condensing
Storage Temperature
% RH
-10°F to 150ºF (-20 to 65ºC)
0% to 95% non-condensing
Inlet Side Outlet Side
Figure 2. Laboratory Room Dual Duct Supply Air Terminal Components.