Installation Instructions
Installation Instructions
Document No. 570-112
June 14, 2013
Retracting Cable Sash Sensor
Item No. 570-112. Rev. AA Page 1 of 3
Product Description
The Retracting Cable Sash Sensors are compact,
economical and water resistant devices with flexible
cables, spring-loaded spools and a potentiometer to
detect and measure linear position.
Product Number
50" Retracting Cable Sash Sensor
546-04000
80" Retracting Cable Sash Sensor
546-04001
Warning/Caution Notation
WARNING
Personal injury/loss of life may
occur if you do not follow the
procedures as specified.
CAUTION
Equipment damage or loss of data
may occur if you do not follow the
procedures as specified.
Required Tools
Small flat blade screwdriver
3/8-inch open end wrench
Needle nose pliers
Prerequisites
For retrofit installations, notify the laboratory
safety officer that you will be working in the fume
hood. Take the necessary precautions to have
the fume hood cleaned, prepared and safety
certified.
For best appearance, mount the sensor so that
the cable passes on the inside of the sash.
The surface on which you mount the sensor
spool should be smooth, flat, clean, and free of
obstacles such as rivets, screw heads, nozzles,
sash stops, and abrasives. Some installations
may require fabrication of additional struts or
brackets, not included.
Mount the sensor on the fume hood structure,
outside of the area with chemical or corrosive
fumes.
Mount in a location accessible for service and
replacement.