Installation Instructions
Table Of Contents
- Figure 1. UniTrak Assembly Kit.
- Figure 2. Cutting UniTrak Sensor Using the Trim Tool Kit.
- Figure 3. Outside Installation of UniTrak Sensor.
- Figure 4. Inside Installation of UniTrak Sensor.
- Figure 5. Removal of Front Access Panel.
- Figure 6. UniTrak Mounting for Kewaunee Visionaire and Kewaunee Airflow Supreme (Combination-Style Fume Hoods).
- Figure 7. UniTrak Mounting for Hamilton Safeair (Combination-Style Fume Hood).
- Figure 8. UniTrak Mounting (Horizontal-Style Fume Hood).
- Figure 9. Location of Vertical Linkage.
- Figure 10. Measuring Between Sash and UniTrak Sensor.
- Figure 11. Marking First Linkage Bend.
- Figure 12. Linkage on 45( Angle.
- Figure 13. Linkage on a 90( Angle.
- Figure 14. Cutting the linkage.
- Figure 15. Attaching Linkage to Actuator Block.
- Figure 16. Location of Horizontal Linkage.
- Figure 17. Linkage Clearance Measurement.
- Figure 18. Linkage Clearance Measurement for Dual UniTrak sensors.
- Figure 19. Marking the First Linkage Bend.
- Figure 20. First Linkage Bend.
- Figure 21. Second Linkage Bend.
- Figure 22. Marking the Final Linkage Bend.
- Figure 23. Completed Horizontal Linkage.
- Figure 24. Linkage Mounting Block Locations (Two and Four sash configurations).
- Figure 25. Linkage mounted inside fume hood with decorator plate.
- Figure 26. FHC Wiring.
- Figure 27. Sash Wire Numbering.
- Figure 28. Routing Wires Through Fume Hood Wall.
- Two Horizontal Sash Panels.
- Four Horizontal Sash Panels.
- Figure 29. Sash Wire Numbering.
- Figure 30. Sash Sensor on Combination Fume Hood.
- Figure 31. Wire terminations for sash 2 and 4 (for Figure 30)
- Figure 32. Bushing Placement for Outside Mounted UniTrak Wiring.
- Figure 33. Sash sensor shield on vertical sensor
- Figure 34. Sash sensor shield on horizontal sensor.
Document No. 546-00449
Installation Instructions
August 10, 2004
ACTUATOR
LENGTH OF SENSOR
TRIM LINE
CUT
UNITRAK
SENSOR
SILICON SEALANT
SEAL END OF SENSOR
WITH THE SILICON INCLUDED
IN THE SMALL WHITE TUBE.
FUM0337R1
Figure 2. Cutting UniTrak Sensor Using the Trim Tool Kit.
3. Use the alcohol towelette to clean the surface on
which the UniTrak sensor is to be applied. Dry
the surface thoroughly before mounting the
UniTrak.
4. Remove the protective backing from the sensor
strip. Make sure the sensor connector is
positioned at the top of the sensor.
5. Be certain that there is a gap of least 3/8-inch
between the sensor strip and the sash panel.
6. When attaching the UniTrak to the mounting
surface, hold one end up and lightly place the
opposite end down. The UniTrak should be
mounted as straight (parallel to the sash) as
possible.
NOTE: If applied without pressure to the
mounting surface, you may remove
and reapply the UniTrak if
necessary. If pressure is applied, the
adhesive is permanent.
7. Locate the correct mounting area for the UniTrak
sensor. See Figure 3 and Figure 4.
8. Slide the actuator block to the bottom of the
UniTrak sensor strip.
Figure 3. Outside Installation of UniTrak Sensor.
Figure 4. Inside Installation of UniTrak Sensor.
Page 4 of 13 Siemens Building Technologies, Inc.