Data Sheet for Product

Unrestricted
Page 4 of 12
Siemens Industry, Inc.
Lab DXR BACnet MSTP ***
Control Package
Includes the following Control Components:
Actuator Part Number
Transducer Part Number
Flow Transmitter Part
Number
Controller Part Number
EO10
GNP191.1P
DXA.S04P1
MO10
GMA 151.1P
DXA.S04P1
MX10
GDE 161.1P
DXA.S04P1
MX13
GDE 161.1P
DXA.S04P1
DXR2.M17C-103B
EO13
GNP191.1P
DXA.S04P1
DXR2.M17C-103B
*** MSTP control package assemblies currently only orderable through Lab Custom department
Control Components
Part Number
Description
Part Number
Description
540-380
Auto-Zero Module
546-00705
Variable Volume Fume Hood Controller,
Application 941 and 942
540-104
Constant Volume TEC with Auto-
Zero Module
570-00701
DXR2.E17C-103B
DXR2.M17C-103B
550-819B
DXA.S04P1
590-780
GDE161.1P
BACnet Fume Hood Controller,
Applications 6740, 6741 and 6742
Lab DXR BACnet IP Controller, 30 dp
Lab DXR BACnet MSTP controller, 30 dp
BACnet OAM - Off-board Air Module
Lab DXR Airflow Pressure Sensor 0-1”
Differential Pressure Transmitter, 1” WC,
4-20 mA, 0.4% accuracy
Fail-in-Last Position, Modulation, 44 in-lb
electric actuator
GDE131.1P
Fail-in-Last Position, Floating,
44 in-lb electric actuator
GDE161.1P
Fail-safe Spring Return Modulating 62 in-
lb electric actuator
GMA131.1P
Fail-safe Spring Return Floating,
62 in-lb electric actuator
GNP191.1P
GNP Fast Acting Lab Electronic Actuator
3. Choose R for the Mounting Side and append the letter to the part number. (The R is required.)
4. Choose the Inlet size (the size of the duct) and append the 2-digit number to the part number.
Inlet Size
(in inches)
2-digit Number
Inlet Size
(in inches)
2-digit number
4
04
11
11 (12” casing with special orifice)
6
06
12
12
7
07
14
14
8
08
16
16
9
09
18
18
10
10
5. Choose the Casing Material and the Sensor, and append the letters to the part number:
Casing Material and Sensor Type
AA
Galvanized steel casing with orifice sensor.
AB
Galvanized steel casing with multi-point sensor. (Does not apply to inlet sizes 4, 11, or 18.)
BA
Stainless steel casing with orifice sensor.
CA
Teflon®-coated steel casing with orifice sensor.
6. Select the End fitting, and append that letter to the part number: