Basic Documentation
Table Of Contents
Product or company names mentioned herein may be the trademarks of their respective owners. © 2009 Siemens Industry, Inc.
Siemens Industry, Inc. Printed in the USA
Building Technologies Division Document No. 149-997
1000 Deerfield Parkway Page 10 of 10
Buffalo Grove, IL 60089-4513
USA
corridor. Exhaust air from the laboratory suite,
laboratory rooms, fume hoods and biological safety
cabinets must not be re-circulated to other rooms or
parts of the facility, even when HEPA filtered. As
with chemical fume exhaust, the air exhausted from
biological laboratories must be discharged outside of
the building and dispersed in a manner that does not
pose a hazard to persons, other buildings, or allows
entrainment into building fresh air intakes.
Isolation Dampers
Biosafety cabinets are periodically decontaminated
using formaldehyde or vaporized hydrogen peroxide
while the laboratory room is unoccupied. If a spill or
release of some biological agent occurs within the
laboratory, the room itself may also need to be
decontaminated. The BMBL requires that BSL-3
laboratories be designed to accommodate these
procedures. Therefore, a bubble tight isolation
damper may be located between the room supply air
terminal and room supply air diffuser(s). Exhaust air
bubble tight isolation dampers should be located
after the juncture of all separate room exhaust
provisions and the final HEPA filters associated with
the room exhaust. If multiple laboratories have their
exhausts combined into a single manifold type
exhaust system, each individual laboratory room
exhaust must be capable of being isolated from the
other room exhausts by a bubble tight damper.
Access to such dampers, as well as all duct
components, should preferably be from outside of
the laboratory suite.
Maintenance and Housekeeping
Perhaps one of the most overlooked aspects of safe
and efficient biological laboratory operations is that
of maintenance. As with any type of facility,
maintenance tasks must be routinely performed.
This includes servicing laboratory equipment as well
as elements of the facility itself such as HVAC
components, plumbing fixtures, electrical light
fixtures, etc. Those performing such tasks must
receive proper training regarding the potential
hazards of the laboratory environment along with the
proper entry and exit procedures.
Laboratory suites should be designed to ensure that
maintenance personnel can only enter and exit by
the normal process. This includes clothes changing,
gowning, etc. In the case of maintenance workers,
additional problems can exist with respect to the
need to have tools and other equipment such as
ladders, portable lights, etc. If these items are
brought into a BSL-3 laboratory or support room,
there must be provision to ensure adequate
decontamination before being removed from the
laboratory suite. Since this may not be easily
accomplished, it is advantageous to provide a
storage provision within the laboratory suite for such
items so that workers do not have to bring in
commonly needed tools and equipment every time a
maintenance task needs to be done.
Housekeeping tasks are normally performed by
laboratory assistant personnel who are aware of the
potential hazards and safety procedures necessary
to protect against accidental contact with biological
agents, aerosols, and especially how to prevent the
potential migration of such agents outside of the
laboratory. It is recommended that adequate
housekeeping storage also be provided within the
laboratory suite for items normally needed for
housekeeping purposes. Items include cleaning
supplies, mops, janitorial sinks, disinfectants, wipes,
tissues, etc. Note that Figure 1 and Figure 5 show
Maintena
nce and Janitorial Closets within the
laboratory suite for this purpose.
In addition, there must be a procedure for the safe
removal and disposal of regularly discarded items
from within the laboratory suite, such as waste
paper, laboratory wipes, broken glassware, and
other refuse.