Data Sheet for Product
6
Siemens Industry, Inc.
A6V11402065_en
Smart Infrastructure
2020-07-30
Technical data
Housing
Dimensions 215.0mm x 160.0mm x 16.51mm
(9.25” W x 6.3” H x 0.65”D215)
Weight 0.9 kg (2 lbs).
Mounting Triple-gang, double deep electrical box. RACO 697,
Appleton M3-350, or equivalent.
Case Fire-retardant plastic UL94 V-0.
Display 7” Projected Capacitive (PCAP) multitouch. 800 x 480 pixels.
Usable with medical gloves.
Display brightness 1…7
Audible alarm Dual piezo with 3 volume levels (0-75 dB max.)
Pressure sensor specifications*
Accuracy RSS ±0.25%
Non-linearity (BFSL) ±0.24%
Hysteresis ±0.05%
Non-repeatability ±0.05%
Span setting tol. ±0.5% Rdg
Zero/span shift % FS 206.8 Pa (± 0.03% PSI)
Overpressure 6894.8 Pa (±1 PSI or 15” WC for ≤0.10” WC FS)
Pressure media Air or non-conductive, non-explosive gases
Pressure fittings 3/16” barbed fittings
Altitude 2000 m (6562 ft.) max.
Position Housing to be 90° in reference to level surface, ±5°
* These specifications apply to all sensor ranges.
Power supply
Operating voltage AC 24 V
Frequency 50/60 Hz
Power consumption including connected field devices AC 18…32 V
Power draw 17 W max.
Inputs
Voltage measurement, analog
Type
Range
Resolution
Accuracy
AI 0...10 V 0...10 V (-1...11 V) 2 mV 10 mV / 0.1% FS
AI 4…20 mA * 4…20 mA (-1...11 V) .01mA (13 μA) 20 mA / 0.1% FS
* For 4-20 mA loops, it is required that a 250 ohm resistor be field-installed between the terminals labeled UI/AI and
COM. The Room Condition Monitor measures the voltage across this resistor as 1…5 VDC corresponding to
4…20 mA.