Commissioning Instructions
Table Of Contents
- Before You Begin
- Verifying Power
- Verifying Slave Mode Application Number
- Setting Controller Address
- Setting the Application
- Setting Display Units
- Testing the Operator Display Panel
- Setting Duct Area
- Setting Airflow Sensing Input
- Setting Flow Coefficients
- Automatic Calibration Option
- Setting Blank Display
- Setting Display Weight
- Setting Display Resolution
- Changing Exhaust Minimum
- Changing Exhaust Maximum
- Changing Face Velocity Setpoints and OCC Delay
- Setting Hi and Low Warn Limits
- Setting Hi and Low Alarm Limits
- Setting Alarm Timer
- Setting Emergency Setpoint
- Setting Emergency Timer
- Setting Remote Purge
- Fume Hood Specific Sash Setup and Calibration
- Setting External Face Area Input
- Setting Sash Area Alarms
- (Optional) Setting Damper Control — Application 2941
- Checkout of Damper — Application 2941
- (Optional) Setting Airflow Input Type
- (Optional) Calibrating the DP Transmitter without an Autozero Module
- (Optional) Calibrating with an Autozero Module
- AVS FAILMODE
- Setting Airflow Control — Application 2942
- Range of Airflow Control — Application 2942
- Configuring Airflow Control — Application 2942
- Setting AO2 Range
- Setting AO2 Voltage Minimum
- Start-up/Decommission Mode
- Loop Tuning Procedures
- Flashing Controller Firmware
Before You Begin
Fume Hood Specific Sash Setup and Calibration
15
Siemens Industry, Inc. Start-up Procedures 140-1321
Restricted 2015-11-04
NOTE:
The fixed area of the fume hood is an area that remains open
regardless of sash position or movement. For example, most fume hoods
have an intake gap under the lower airfoil and above the cabinet of the fume
hood (typically a 1 inch gap). Also include 1% of the maximum open face area
in this calculation for other open areas, such as the space between the sash
and the track, and leakage.
Vertical Sash Fume Hood with Fixed Area.
8.
Set CAL SASH NUM to
1
.
9.
Set CAL SASH LOC to
MIN
for minimum opening.
10.
Close the sash door.
11.
Set CAL SASH POS to
0
inches (minimum) unless a physical stop prevents
the sash from fully closing. If a physical stop is present, measure the distance
from the lower end of the vertical track to the bottom of the sash in inches
(cm). Set CAL SASH POS to this value.
12.
Set CAL SASH LOC to
MAX
for maximum opening.
13.
Open the sash to the fully opened position.
14.
Measure the distance from the lower end of the vertical track to the bottom
edge of the sash in inches (cm). Set CAL SASH POS to this value
(maximum).
15.
Set CAL SASH NUM to
0
to turn the calibration sequence OFF.