98 Siemens-Elema AB E382 E380E 061 01 03 01
Disassembling and assembling Servo Ventilator 300/300A
Disassembling the pneumatic
section
Inspiratory valves (gas modules)
Remove the screw (1).
Push down the safety catch (2).
Pull out the inspiratory valve (3).
If the inspiratory valve unit – Optional (4)
is replaced with units for SUN 345 or NO
gas administration, these units must also
be removed.
The inspiratory valves are factory
calibrated. They must not be disassem-
bled further than described in the
Operating Manual, chapter ”3000 hours
overhaul with complete cleaning”.
Pneumatic section casing
Remove the inspiratory and expiratory
channels as described in the Operating
Manual, chapter ”3000-hours overhaul
with complete cleaning”.
Remove the plastic cover (1). This cover is
not included in newer units.
Pull out and turn the locking pins (2) from
the locked position.
Lift the casing (3) carefully until the fan
connector (4) is accessible.
Disconnect the fan connector (4).
Lift off the casing (3).
300-A41X
3
2
4
1
2
3
1
4
300-A42X