Service manual

3
Servo Ventilator 300/300A Disassembling and assembling
E382 E380E 061 01 03 01 Siemens-Elema AB 99
Cover plate
Remove the inspiratory valves for AIR and
O
2
as previously described.
Remove the pneumatic section casing as
previously described.
Disconnect the flow transducer cable (1).
Disconnect the O
2
cell cable (2).
Remove the three screws (3).
Carefully lift off the cover plate (4).
Expiratory valve
Remove the cover plate as previously
described.
Disconnect the two expiratory valve
connectors (1 and 2).
Remove the three screws (3).
Lift off the expiratory valve (4).
2
1
4
3
300-A44X
300-A43X
4
3
1
2