Service procedures Servo Ventilator 300/300A
Replacing the fuses
Fuses F1, F2, F3 and F4
Fuses F1 – F4 are mounted on PC 1618
POWER SUPPLY inside the power section.
Disconnect and prepare the ventilator as
described in chapter ”3. Disassembling
and assembling”, section Preparation”.
Disassemble the power section as
described in chapter ”3. Disassembling
and assembling”, section Disassembling
the power section”.
Replace blown fuses with new fuses as
listed in the table below:
Fuses Purpose
F1: F 6.3 A Limits the total +24 V supply current
F2: F 1.6 A Controls the +24 V FUSED voltage
F3: F 6.3 A Internal battery connector fuse
F4: F 6.3 A External battery connector fuse
Assemble the power section.
Fuses F11 and F12
Disconnect the mains power cable from
the ventilator.
Open the cover (1).
Pull out the fuse holders (2).
Replace blown fuses with new fuses as
listed in the table below:
Mains power Selector Fuses
voltage setting F11/F12
100 V (±10%) 100 V T 3.15 AL
120 V (±10%) 120 V T 3.15 AL
220 V (±10%) 220 V T 1.6 AL
230 V (±10%) 220 V T 1.6 AL
240 V (±10%) 240 V T 1.6 AL
Close the cover (1). The selected voltage
is shown in the cover window.
1
F11
F12
300-L28X
2