Service manual

26 Siemens-Elema AB E382 E380E 061 01 03 01
Description of functions Servo Ventilator 300/300A
2
1.3 Respiratory pattern
For control and/or monitoring of respiratory
pattern parameters:
4 four-character LED displays
5 potentiometer knobs:
CMV freq. b/min. This control is
equipped with two parallell
potentiometers – one for the control
function and the other for the monitor/
alarm function.
Insp. time %. Push button release
function for two-way mechanical stops
at approx. 20, 50 and 70%.
Pause time %. Push button release
function for one-way mechanical stop at
approx. 20%.
Insp. rise time %. Push button release
function for one-way mechanical stop at
approx. 1%.
SIMV freq. b/min.
5 yellow indicator LEDs.
1.4 Volume
For control and/or monitoring of minute
volume and tidal volume parameters:
2 four-character green LED displays.
Showing information generated via
2.1 PANEL CPU.
3 four-character red LED displays.
Showing information generated via 4.1
MONITOR/ALARM CPU.
2 fields of double column LED bargraph.
3 potentiometer knobs:
Volume
Upper alarm limit
Lower alarm limit.
3 yellow indicator LEDs.
1.2 Airway pressure
For control and monitoring of AIRWAY
PRESSURE
parameters:
4 four-character LED displays.
1 double column LED bargraph.
5 potentiometer knobs:
Upper press. limit. Push button release
function for one-way mechanical stops
at approx. 60, 80 and 100 cm H
2
O.
Pressure Control Level above PEEP.
Push button release function for one-
way mechanical stops at approx. 30, 60
and 80 cm H
2
O.
Pressure Support Level above PEEP.
Push button release function for one-
way mechanical stops at approx. 30, 60
and 80 cm H
2
O.
PEEP. Push button release function for
one-way mechanical stop at approx.
20 cm H
2
O.
Trigg. sensitivity Level below PEEP.
5 yellow indicator LEDs.