Service manual
48 Siemens-Elema AB E382 E380E 061 01 03 01
Description of functions Servo Ventilator 300/300A
2
5 Optional PC board slot
Not used in this version.
4.6 Barometer pressure
transducer
Pressure sensor measuring the ambient
pressure. The signal is used in the block
8.11 PRESSURE CALCULATOR for the regulation
of inspiratory flow in the main block 8
INSPIRATORY VALVE UNIT – AIR as well as in the
main block 9 INSPIRATORY VALVE UNIT – O
2
.
The measured value can be read in the
Alarm and messages display on the front
panel, see chapter Calibration in the SV 300
– Operating Manual.
4.7 Loudspeaker
For audible alarm generated by the regular
(digital) monitor/alarm block 4.1 MONITOR/
ALARM CPU.
4.8 Beeper
Tone generator for audible alarm controlled
by the (analog) block 4.5 BACKUP ALARM
SYSTEM
.
4.9 Failure alarm box (FAB)
A FAILURE ALARM BOX (FAB) must be installed on
SV 300 equipped with MON-PROM V5.01. The
alarm functions of the FAB are integrated on
MON-PROM V6.00/PC 1608E and the FAB is not
used on SV 300 equipped with MON-PROM
V6.00/PC 1608E (or higher). See section
”4 MONITORING” for further information
regarding PROM and PC board versions.
The FAB is mounted on the SV 300
connector N81. All functionality of the SV
300 Analog
I/O terminal N81 is available through the
one-to-one connected FAB connector N81.
The FAB includes the functional blocks:
+24 V MONITOR
FAB INTERNAL MONITOR
ALARM CONTROL
A FAB BEEPER, a red LED, a push button and
the connector N81 (as described above) are
also included in the FAB.
The FAB is designed to detect loss of the
SV 300 internal +24 V power and emits
audible alarm in response to such a power
failure or in case of a FAB internal failure.
For a further functional description of the
FAB, contact your Siemens representative.