Service manual

70 Siemens-Elema AB E382 E380E 061 01 03 01
Description of functions Servo Ventilator 300/300A
2
12 Inspiratory pressure
The function of this main block is to
measure the pressure in the 11 INSPIRATORY
CHANNEL
and generate a corresponding signal
INSP. PRESSURE via the board PC 1607
PNEUMATIC INTERCONNECTION (N24:C7) to the
main blocks 4 MONITORING and 6 REFERENCE &
TIMING.
12.1 Bacteria filter
This filter protects the 12.2 PRESSURE
TRANSDUCER
on the board 12.3 PRESSURE
AMPLIFIER
from contamination.
Service interval 3000 h of operation for
exchange of the inspiratory pressure BACTERIA
FILTER
.
12.2 Pressure transducer
This differential PRESSURE TRANSDUCER is an
integrated part of the board 12.3 PRESSURE
AMPLIFIER
mounted together with a silicon
rubber connection piece. It refers the
pressure to ambient pressure and gives a
linear measurement in the range –20 cm H
2
O
to +120 cm H
2
O.
Pressure exceeding ±200 cm H
2
O must be
avoided.
12.3 Pressure amplifier
This board PC 1611 contains:
a pressure transducer
a generator of +5V REFERENCE VOLTAGE
an amplifier for the pressure signal
the two potentiometers for calibration of
zero offset and gain.
When calibrated according to the Operating
Manual, chapter Calibration, the PRESSURE OUT
signal (P24:C7) will have a scale factor of
75 mV/cm H
2
O.