Servo Ventilator 300/300A Disassembling and assembling
E382 E380E 061 01 03 01 Siemens-Elema AB 93
Separating the pneumatic
section from the power
section
Place the patient unit on the edge of the
workbench so the screws are accessible
from below.
Loosen the four screws (1).
Lift off the pneumatic section from the
power section. Place the pneumatic
section on a support beside the power
section.
Disconnect the battery connector (2).
To reduce the risk of damages, caused
by short-circuits in the power section,
always have the batteries disconnected
when the power section is open.
Instead of cutting the cable ties, remove
the screw (3) holding the battery
bracket (4).
Disconnect the connectors (5, 6 and 7).
The pneumatic section is now separated
from the power section.
1
5
6
7
3
4
300-A35X
2