Service manual

3
94 Siemens-Elema AB E382 E380E 061 01 03 01
Disassembling and assembling Servo Ventilator 300/300A
Disassembling the power
section
Mains power inlet
Separate the pneumatic section from the
power section as previously described.
Remove the cover (1).
On older units: A small plastic cover
mounted with two screws (2).
On newer units: A big metal cover (as
illustrated) mounted with three screws (3).
An EMI-filter is attached to the cover with
a cable tie. Cut the cable tie if the cover
must be completely removed from the
power section.
Disconnect the cables from the mains
power inlet.
Remove the mains power inlet (4) from
the power section.
Transformer
Separate the pneumatic section from the
power section as previously described.
Remove the cover (1) from the mains
power inlet as previously described.
Disconnect the transformer cables from
the mains power inlet.
Disconnect the transformer cables from
the capacitor (2) and from the screw
terminal (3).
Remove the nut (4), washer (5) and rubber
plate (6).
Lift off the transformer (7).
3
4
5
6
7
1
2
300-A37X
3
4
1
300-A36X
2