Service manual

3
Servo Ventilator 300/300A Disassembling and assembling
E382 E380E 061 01 03 01 Siemens-Elema AB 95
Power supply board PC 1618
Separate the pneumatic section from the
power section as previously described.
Remove the cover (1) from the mains
power inlet.
Remove the four screws (2).
Disconnect the;
Cables at screw terminal (3).
Ground cable (4). This ground cable is
used only in older versions. It has been
removed in newer versions.
External battery inlet connector (5).
Operating time meter connector (6).
Remove the nuts holding the diodes (7)
and release the diodes from the screws.
Remove the nuts and screws holding the
two auxiliary equipment outputs (8).
Remove the screw (9).
Remove the capacitor holder (10).
Carefully lift off the PC 1618.
2
300-A38X
2
3
4
5
6 7
8
9
10
1