Introduction 1 Safety information 2 SITRANS L Description 3 Radar Transmitters SITRANS LR250 with mA/HART Installing/mounting 4 Connecting 5 Commissioning 6 Remote operation 7 Parameter reference 8 Service and maintenance 9 Operating Instructions 7ML5431 (LR250) 7ML5432 (LR250 flanged encapsulated antenna) 7ML5433 (LR250 hygienic encapsulated antenna) 10/2019 A5E32220602-AG Diagnosing and troubleshooting 10 Technical data 11 Dimension drawings 12 Appendix A: Technical reference
Legal information Warning notice system This manual contains notices you have to observe in order to ensure your personal safety, as well as to prevent damage to property. The notices referring to your personal safety are highlighted in the manual by a safety alert symbol, notices referring only to property damage have no safety alert symbol. These notices shown below are graded according to the degree of danger.
Table of contents 1 2 3 4 Introduction...................................................................................................................................................9 1.1 LR250 mA/HART manual usage..............................................................................................9 1.2 Purpose of this documentation.................................................................................................9 1.3 Revision history......................................
Table of contents 5 6 7 4.3.3 4.3.4 4.3.5 Orientation in a vessel with obstructions................................................................................29 Mounting on a Stillpipe or Bypass Pipe .................................................................................29 Device orientation ..................................................................................................................30 4.4 4.4.1 4.4.2 4.4.3 Installation instructions.................................
Table of contents 8 7.1.4 7.1.5 7.1.6 7.1.6.1 7.1.6.2 7.1.6.3 7.1.6.4 7.1.6.5 7.1.6.6 7.1.6.7 7.1.6.8 7.1.6.9 7.1.6.10 7.1.6.11 7.1.6.12 7.1.6.13 7.1.6.14 7.1.6.15 7.1.6.16 7.1.6.17 7.1.6.18 7.1.6.19 Quick Start Wizard via SIMATIC PDM ...................................................................................66 Changing parameter settings using SIMATIC PDM...............................................................70 Parameters accessed via pull-down menus......................................
Table of contents 10 11 12 6 9.2 Cleaning ...............................................................................................................................153 9.3 9.3.1 9.3.2 Maintenance and repair work...............................................................................................154 Unit repair and excluded liability ..........................................................................................155 Part replacement............................................
Table of contents A 12.15 Hygienic encapsulated antenna (DN80 nozzle/slotted nut to DIN 11851) ..........................198 12.16 Hygienic encapsulated antenna (DN100 nozzle/slotted nut to DIN 11851) ........................199 12.17 Hygienic encapsulated antenna (DN50 aseptic slotted nut to DIN 11864-1) ......................200 12.18 Hygienic encapsulated antenna (DN80 aseptic slotted nut to DIN 11864-1) ......................201 12.
Table of contents B C A.5.1 A.5.2 A.5.3 A.5.4 A.5.5 A.5.6 A.5.7 Pressure Equipment Directive, PED, 14/68/EU ...................................................................235 Stainless steel horn antenna................................................................................................235 Stainless steel flanged horn antenna ...................................................................................236 PTFE flanged encapsulated antenna............................................
1 Introduction 1.1 LR250 mA/HART manual usage Note This manual applies to the SITRANS LR250 mA/HART version only. HART® is a registered trademark of the HART Communication Foundation. Follow these operating instructions for quick, trouble-free installation, and maximum accuracy and reliability of your device. We always welcome suggestions and comments about manual content, design, and accessibility. Please direct your comments to: Technical publications (mailto:techpubs.smpi@siemens.com) 1.
Introduction 1.4 Industrial use note Edition Firmware rev. PDM EDD rev. Date Changes 3.1 1.02.01 2 Apr 2009 ● AMS EDD Rev. 1.02.01 1.02.01 ● Support NAMUR NE 43 ● Harmonization of menu structures and parameter names across products ● Display indicates progress towards first measurement 3.2 1.02.03 1.02.01 16 June 2010 ● Display contrast improvement ● Antenna type parameter cannot be modified 4.0 1.03.02 (re‐ quires HW 2.0.0) 1.02.01 16 June 2010 ● Low current HW 2.0.0 supported 4.0 1.03.
Introduction 1.6 Security information 1.5 Checking the consignment 1. Check the packaging and the delivered items for visible damages. 2. Report any claims for damages immediately to the shipping company. 3. Retain damaged parts for clarification. 4. Check the scope of delivery by comparing your order to the shipping documents for correctness and completeness. WARNING Using a damaged or incomplete device Risk of explosion in hazardous areas. ● Do not use damaged or incomplete devices. 1.
Introduction 1.8 Notes on warranty 1.7 Transportation and storage To guarantee sufficient protection during transport and storage, observe the following: ● Keep the original packaging for subsequent transportation. ● Devices/replacement parts should be returned in their original packaging. ● If the original packaging is no longer available, ensure that all shipments are properly packaged to provide sufficient protection during transport.
2 Safety information 2.1 Preconditions for safe use This device left the factory in good working condition. In order to maintain this status and to ensure safe operation of the device, observe these instructions and all the specifications relevant to safety. Observe the information and symbols on the device. Do not remove any information or symbols from the device. Always keep the information and symbols in a completely legible state. 2.1.
Safety information 2.1 Preconditions for safe use 2.1.3 FCC Conformity US Installations only: Federal Communications Commission (FCC) rules Note ● This device has been tested and found to comply with the limits Class B digital device part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. ● This device has also been tested and found to comply with the limits §15.
Safety information 2.1 Preconditions for safe use Note ● This device has been tested and found to comply with the limits RS- 211 - Level Probing Radar Equipment. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. ● This device is certified to measure levels in fixed or mobile enclosed tanks made from metal, concrete or materials with similar RF attenuating properties.
Safety information 2.1 Preconditions for safe use 2.1.6 Radio Equipment Directive (RED) 2014/53/EU For the receiver test that covers the influence of an interferer signal to the device, the performance criterion has at least the following level of performance according to ETSI TS 103 361 [6]: ● Performance criterion: measurement value variation Δd over time during a distance measurement ● Level of performance: Δd ≤ ±50 mm 2.1.
Safety information 2.4 Use in hazardous areas 2.2 Improper device modifications WARNING Improper device modifications Risk to personnel, system and environment can result from modifications to the device, particularly in hazardous areas. ● Only carry out modifications that are described in the instructions for the device. Failure to observe this requirement cancels the manufacturer's warranty and the product approvals. 2.
Safety information 2.4 Use in hazardous areas WARNING Loss of safety of device with type of protection "Intrinsic safety Ex i" If the device has already been operated in non-intrinsically safe circuits or the electrical specifications have not been observed, the safety of the device is no longer ensured for use in hazardous areas. There is a danger of explosion. ● Connect the device with type of protection "Intrinsic safety" solely to an intrinsically safe circuit.
3 Description 3.1 SITRANS LR250 overview WARNING Loss of protection Danger to personnel, system and environment can result from improper use of the device. ● SITRANS LR250 is to be used only in the manner outlined in this manual, otherwise protection provided by the device may be impaired. SITRANS LR250 is a 2-wire 25 GHz pulse radar level transmitter for continuous monitoring of liquids and slurries in storage vessels including high pressure and high temperature, to a range of 20 meters (66 feet).
Description 3.4 Approvals and certificates 3.2 Programming This device is very easy to install and configure via a graphical local user interface (LUI). You can modify the built in parameters either locally via the Siemens infrared handheld programmer, or from a remote location using one of the following options: ● SIMATIC PDM ● AMS Device Manager ● FDT/DTM platform (such as PACTware™ or FieldCare) ● HART Handheld 375/475 3.
Installing/mounting 4.1 4 Basic safety information Note Material compatibility Siemens can provide you with support concerning selection of sensor components wetted by process media. However, you are responsible for the selection of components. Siemens accepts no liability for faults or failures resulting from incompatible materials. WARNING Unsuitable connecting parts Risk of injury or poisoning. In case of improper mounting, hot, toxic, and corrosive process media could be released at the connections.
Installing/mounting 4.1 Basic safety information 4.1.1 Unsuitable cables, cable glands and/or plugs WARNING Unsuitable cables, cable glands and/or plugs Risk of explosion in hazardous areas. ● Use only cable glands/plugs that comply with the requirements for the relevant type of protection. ● Tighten the cable glands in accordance with the torques specified in Installation instructions (Page 30). ● Close unused cable inlets for the electrical connections.
Installing/mounting 4.1 Basic safety information WARNING Exceeded maximum permissible operating pressure Danger of injury or poisoning. The maximum permissible operating pressure depends on the device version. The device can be damaged if the operating pressure is exceeded. Hot, toxic and corrosive process media could be released. ● Make sure that the device is suitable for the maximum permissible operating pressure of your system.
Installing/mounting 4.2 Installation location requirements 4.2 Installation location requirements WARNING Aggressive atmospheres Danger to personnel, system and environment can result from unsuitable environment. ● Provide an environment suitable to the housing rating and materials of construction. NOTICE Direct sunlight Device damage. The device can overheat or materials become brittle due to UV exposure. ● Protect the device from direct sunlight.
Installing/mounting 4.3 Proper mounting Hygienic encapsulated ‑40 to +80 °C (‑40 to +176 °F) ‑40 to +170 °C (‑40 to +338 °F) with FKM seals used on process connection: -20 to +170 °C (-4 to +338 °F) with EPDM seals used on process connection: -40 to +120 °C (-40 to +248 °F) Note Details about the process connection, process temperature and materials are laser etched into the body of the flanged and hygienic versions. All other SITRANS LR250 versions have details listed on a tag. 4.
Installing/mounting 4.3 Proper mounting 4.3.1 Nozzle design Threaded PVDF antenna Stainless steel horn antenna Polymeric lens versions Hygienic encapsulated antenna (HEA) ① ② ③ ④ Minimum clearance: 10 mm (0.4") Minimum diameter: 50 mm (2") Maximum nozzle length Maximum length/diameter ratio 1:1 ● The end of the antenna must protrude a minimum of 10 mm (0.4") to avoid false echoes being reflected from the nozzle1).
Installing/mounting 4.3 Proper mounting Preferred Undesirable Beam angle Note ● Beam width depends on antenna size and is approximate: see below. ● For details on avoiding false echoes, see Auto False Echo Suppression (Page 228). ● Beam angle is the width of the cone where the energy density is half of the peak energy density. ● The peak energy density is directly in front of and in line with the antenna. ● There is a signal transmitted outside the beam angle, therefore false targets may be detected.
Installing/mounting 4.3 Proper mounting Antenna type Antenna size PTFE flanged encapsulated 2" Class 150 ASME B16.5 12.8° 3, 4, 6" Class 150 ASME B16.5 9.6° 50A 10K JIS B 2220 12.8° 80A/100A/150A 10K JIS B 2220 9.6° DN50 PN10/16 EN1092-1 12.8° DN80/DN100/ DN150 PN10/16 EN1092-1 9.6° 2" Sanitary Clamp according to ISO 2852 12.
Installing/mounting 4.3 Proper mounting 4.3.3 Orientation in a vessel with obstructions Polarization reference point For best results on a vessel with obstructions, or a stillpipe with openings, orient the front or back of the device toward the obstructions. For an illustration, see Device orientation (Page 30). 1 2 3 ① ② ③ 4.3.
Installing/mounting 4.4 Installation instructions 1) Mounting in a pipe greater than 100 mm (4") can cause large errors, and therefore is not recommended. Suitable pipe diameters: 40 to 100 mm (1.5 to 4") PVDF antenna 50 mm (2") only Flanged encapsulated antenna 50 to 100 mm (2 to 4") Hygienic encapsulated antenna 50 to 100 mm (2 to 4") Not recommended: > 100 mm (4") Bypass vent: Required at the upper end of the bypass 1) 1) 4.3.
Installing/mounting 4.4 Installation instructions WARNING Improper installation Danger to personnel, system and environment can result from improper installation. ● Installation shall only be performed by qualified personnel and in accordance with local governing regulations. NOTICE Device handling Damage to device may result from improper handling. ● Handle the device using the enclosure, not the process connection or tag, to avoid damage.
Installing/mounting 4.4 Installation instructions 1. Before inserting the device into its mounting connection, check to ensure the threads are matching, to avoid damaging them. 2. Simply screw the device into the process connection, and hand tighten, or use a wrench. 4.4.
Installing/mounting 4.4 Installation instructions 1-3-2-4 1 4 2 3 1-5-3-7 2-6-4-8 1 8 2 7 3 6 4 5 Flange bolting instructions: 1. Use cross-pattern sequence as shown. 2. Check uniformity of the flange gap. 3. Apply adjustments by selective tightening if required. 4. Torque incrementally until desired value is reached. 5. Check/re-torque after 4 to 6 hours. Recommendations for flange bolting: ● Check bolts periodically, re-torque as required.
Installing/mounting 4.4 Installation instructions 4.4.3 Hygienic versions WARNING Loss of sanitary approvals Loss of sanitary approvals can result from improper installation/mounting. ● Take special care when installing in hygienic or sanitary applications. Comply with installation/mounting guidelines to ensure cleanliness and the ability to keep the wetted parts in a position to be readily cleanable. (See relevant EHEDG/3A documentation - not supplied).
Installing/mounting 4.5 Disassembly 4.5 Disassembly DANGER Pressure applications Danger to personnel, system and environment will result from improper disassembly. ● Never attempt to loosen, remove, or disassemble process connection while vessel contents are under pressure.
Installing/mounting 4.
5 Connecting 5.1 Basic safety information WARNING Missing PE/ground connection Risk of electric shock. Depending on the device version, connect the power supply as follows: ● Power plug: Ensure that the used socket has a PE/ground conductor connection. Check that the PE/ground conductor connection of the socket and power plug match each other. ● Connecting terminals: Connect the terminals according to the terminal connection diagram. First connect the PE/ground conductor. 5.
Connecting 5.2 Connecting SITRANS LR250 NOTICE Improper cables and conduit ● Separate cables and conduits may be required to conform to standard instrumentation wiring practices or electrical codes. 1 2 4 3 5 ① ② ③ Use a 2 mm Allen key to loosen the lid-lock set screw c) ④ Cable shield Plug (IP68) ⑤ Ground terminal a) b) b) Optional cable gland (or NPT cable entry) a) May be shipped with the device.
Connecting 5.3 Wiring setups for hazardous area installations Connecting HART Typical PLC/mA configuration with HART 4 1 5 3 2 ① ② ③ Power supply SITRANS LR250 R = 250 Ω 6 ④ ⑤ ⑥ Active PLC HART modem HART communicator Note ● Depending on the system design, the power supply may be separate from the PLC, or integral to it.
Connecting 5.3 Wiring setups for hazardous area installations For installation across Zone 0 and Zone 1 boundary: ① ② ③ 5.3.1 Zone 1 Threaded nut acts as partition wall Material: Stainless steel, aluminum, and potting compound (hardness of at least 88 SHORE D; compressive strength of at least 14,000 psi) Zone 0 Intrinsically safe wiring Device nameplate (ATEX/IECEX/INMETRO/RCM) SITRANS LR250 BOX 2 7MLxxxx-xxxxx-xxxx BOX 1 Serial No: GYZ / S1034567 Encl.: NEMA / TYPE 4X, 6, IP67, IP68 Amb.Temp.
Connecting 5.3 Wiring setups for hazardous area installations FM/CSA Intrinsically Safe connection drawing 23650653 (http:// www.automation.siemens.com/bilddb/index.aspx? gridview=view2&objkey=G_FI01_XX_03549&showdetail=true&view=Search) ● For power demands see Curve 1 (General Purpose, Intrinsically Safe, Non-Sparking, Nonincendive) (Page 245). ● For wiring requirements: follow local regulations.
Connecting 5.3 Wiring setups for hazardous area installations The FM/CSA Class 1, Div 2 connection drawing number 23650673 can be downloaded from the Siemens Industry Image Database: FM/CSA Class 1, Div 2 connection drawing number 23650673 (http:// www.automation.siemens.com/bilddb/index.aspx? gridview=view2&objkey=G_FI01_XX_03551&showdetail=true&view=Search) ● For power demands see Curve 1 (General Purpose, Intrinsically Safe, Non-Sparking, Nonincendive) (Page 245). 5.3.
Connecting 5.4 Instructions specific to hazardous area installations Product page (http://www.siemens.com/LR250) Go to Support > Approvals / Certificates. The IECEx certificate listed on the nameplate can be viewed on the IECEx website: IECEx (http://iecex.iec.ch/) Click on Certified Equipment then enter the certificate number IECEx SIR 08.0107X. ● For power demands see Curve 2 (Flameproof, Increased Safety, Explosion-proof) (Page 246). ● For wiring requirements follow local regulations.
Connecting 5.4 Instructions specific to hazardous area installations 3. The equipment may be used with flammable gases and vapors with apparatus group IIC, IIB and IIA and temperature classes T1, T2, T3 and T4. 4. The equipment has a degree of ingress protection of IP67 and a temperature class of T100 °C and may be used with flammable dusts. 5. The equipment is certified for use in an ambient temperature range of –40 °C to +80 °C. 6.
Commissioning 6.1 6 Basic safety information WARNING Loss of explosion protection Risk of explosion when device is not properly commissioned If opening device ● Isolate from power. - or ● Ensure that the atmosphere is explosion-free (hot work permit). Ensure device is properly closed before returning to operation. 6.2 Operating via the handheld programmer The handheld programmer used with this device contains lithium batteries that are nonreplaceable.
Commissioning 6.2 Operating via the handheld programmer 6.2.1 Power up Power up the device. A transition screen showing first the Siemens logo and then the current firmware revision is displayed while the first measurement is being processed. The first time the device is configured, you will be prompted to select a language (English, German, French, or Spanish). To change the language again, see Language (7.). Press Mode 6.2.2 to toggle between Measurement and Program mode.
Commissioning 6.2 Operating via the handheld programmer 6.2.2.1 The LCD display Measurement mode display Normal operation 1 2 3 4 [M] LEVEL 18.91 21.40 rC DATA EXCH.
Commissioning 6.2 Operating via the handheld programmer PROGRAM mode display Navigation view 3 2 1 ① ② ③ 4 SIGNAL PROC.. 2.8.7 ECHO SELECT SAMPLING ECHO QUALITY TVT SETUP 5 ④ ⑤ Item band Menu bar Current item number Current item Current menu ● A visible menu bar indicates the menu list is too long to display all items. ● A band halfway down the menu bar indicates the current item is halfway down the list.
Commissioning 6.2 Operating via the handheld programmer 6.2.2.2 Handheld programmer (Part No. 7ML1930-1BK) The programmer is ordered separately. Key functions in measurement mode Key Function Result Updates the loop current New value is displayed in LCD secondary re‐ gion. Updates internal enclosure temperature reading Updates echo confidence value New value is displayed in LCD secondary re‐ gion.
Commissioning 6.2 Operating via the handheld programmer 6.2.3 Programming Note ● While the device is in PROGRAM mode the output remains fixed and does not respond to changes in the device. ● The device automatically returns to Measurement mode after a period of inactivity in PROGRAM mode (between 15 seconds and 2 minutes, depending on the menu level). Change parameter settings and set operating conditions to suit your specific application.
Commissioning 6.2 Operating via the handheld programmer ① Handheld programmer ② Display ③ Maximum distance: 300 mm (1 ft) 2. Navigating: key functions in Navigation mode Note ● In Navigation mode ARROW keys move to the next menu item in the direction of the arrow. ● For Quick Access to parameters via the handheld programmer, press Home , then enter the menu number, for example: 2.7.1. (Volume).
Commissioning 6.2 Operating via the handheld programmer ① Parameter name ② Parameter number ③ Current selection 4. Changing a numeric value ● Navigate to the desired parameter. ● Press RIGHT arrow to open parameter view. The current value is displayed. ● Press RIGHT arrow again to open Edit mode. The current value is highlighted. ● Key in a new value. ● Press RIGHT arrow to accept it. The LCD returns to parameter view and displays the new selection. 1 2 LOW CALIB. PT. 1.6 20.00 3 LOW CALIB.
Commissioning 6.2 Operating via the handheld programmer Key Name Function Decimal point Numeric ed‐ iting Enters a decimal point. Plus or minus sign Numeric ed‐ iting Changes the sign of the entered value. Numeral Numeric ed‐ iting Enters the corresponding character. to 6.2.3.1 Quick Start Wizard via the handheld programmer Note ● A reset to factory defaults should be performed before running the Quick Start Wizard if the device has been used in a previous application. See Master Reset (4.1.).
Commissioning 6.2 Operating via the handheld programmer Material Selects the appropriate echo processing algorithms for the material [see Position Detect (2.8.4.2.) for more detail]. Options * LIQUID LIQUID LOW DK a) (low dielectric liquid – CLEF algorithm enabled) a) dK < 3.0 Response Rate Sets the reaction speed of the device to measurement changes in the target range. Use a setting just faster than the maximum filling or emptying rate (whichever is greater). Options * Response Rate (1.3.
Commissioning 6.2 Operating via the handheld programmer Options m, cm, mm, ft, in. Default: m Operation Operation Description NO SERVICE Measurement and associated loop current are not updated, and the device defaults to Fail-safe mode a). LEVEL * Distance to material surface referenced from Low Calibration Point SPACE Distance to material surface referenced from High Calibration Point DISTANCE Distance to material surface referenced from Sensor reference point a) See Material Level (2.5.1.
Commissioning 6.2 Operating via the handheld programmer High Calibration Point Distance from Sensor reference point to High Calibration Point: usually process full level. (See Operation for an illustration.) Values Range: 0.00 to 20.00 m Wizard complete Options BACK, CANCEL, FINISH (Display returns to 1.1 Quick Start Wiz menu when Quick Start is successfully completed.) (Finish). Then press LEFT arrow Press DOWN arrow SITRANS LR250 is now ready to operate. 6.2.3.2 to return to Measurement mode.
Commissioning 6.3 Application examples ① ② ③ ④ ⑤ ⑥ Confidence Distance from Low Calibration Point to vertical cross-hair Algorithm: tF (trueFirst) Distance from flange face to target Pan left/right - selected Pan up/down ● Use UP or DOWN becomes active. ⑦ ⑧ ⑨ ⑩ ⑪ Zoom Measure Exit Exit icon selected Exit icon deselected arrow to scroll to an icon. When an icon is highlighted, that feature ● To move a cross-hair, press RIGHT decrease.
Commissioning 6.3 Application examples To obtain level measurement/4 to 20 mA output proportional to resin levels: ● Low Calibration Pt. = 5 m (16.5 ft) from sensor reference point ● High Calibration Pt.= 0.5 m (1.64 ft) from sensor reference point ● Max.fill/empty rate = 0.2 m/min (0.65 ft/min) In the event of a loss of echo: ● SITRANS LR250 is to go into Fail-safe High after 2 minutes. 1 2 3 4 5 7 6 ① ② ③ ④ Sensor reference point High calibration point 0.
Commissioning 6.3 Application examples 6.3.2 Horizontal vessel with volume measurement Note The minimum distance from the flange face to the target is limited by Near Range (2.8.1.). To obtain level measurement/4 to 20 mA output proportional to vessel volume in a chemical vessel: ● Low Calibration Point = 3.5 m (11.48 ft) from sensor reference point ● High Calibration Point = 0.5 m (1.64 ft) from sensor reference point ● Max. fill/empty rate = 0.2 m/min (0.
Commissioning 6.3 Application examples Parameter type Parameter No./Name Options/ Values Function Independent pa‐ rameters Vessel Shape (2.7.1.1.) PARABOLIC ENDS Defines vessel shape. Maximum Volume (2.7.1.2.) 8000 8000 liters Vessel Dimension A (2.7.1.3.) 0.8 0.8 m (2.62 ft) Vessel Dimension L (2.7.1.4.) 6 6 m (19.68 ft) LOE Timer (2.5.2.) 2 2 minutes Material Level (2.5.1.) HI Fail-safe level set to High Return to Measurement: press Mode 6.3.3 to start normal operation.
Commissioning 6.3 Application examples Parameter type Parameter No./Name Options/Values Function Quick Start Wizard Introduction NEXT Continue with Wizard. Language NEXT Continue with current language. Material LIQUID LOW DK Response Rate MED Medium =1 m/minute Units M meters Operation LEVEL Level is reported as Vol‐ ume when a vessel shape is selected. Low Calibration Point 5 5 m (16.5 ft) High Calibration Point 0.5 0.5 m (1.
Commissioning 6.3 Application examples Propagation Factor 0.9844 0.988 0.9935 0.9965 CLEF Range (2.8.4.4.) settings Low calibra‐ tion point 700 mm (2.29 ft)b) Low calibra‐ tion point 700 mm (2.29 ft)b) Low calibration point -1000 mm (3.28 ft)b) Low calibration point -1000 mm (3.28 ft)b) a) Since pipe dimensions may vary slightly, the propagation factor may also vary.
7 Remote operation 7.1 Operating via SIMATIC PDM SIMATIC PDM is a software package used to commission and maintain process devices. Please consult the operating instructions or online help for details on using SIMATIC PDM. You can find more information on versions and compatibility at our website. See also SIMATIC PDM (www.siemens.com/simatic-pdm) 7.1.1 Functions in SIMATIC PDM Note ● For a complete list of parameters, see Parameter Reference (Page 107).
Remote operation 7.1 Operating via SIMATIC PDM 7.1.1.3 Feature Function Auto False Echo Suppression (Page 74) Screen out false echoes TVT Shaper (Page 73) Manual TVT adjustment Process Variables (Page 79) Monitor process variables and level trend Security (Page 82) Protect security and communication parameters from modification by the maintenance user Features of SIMATIC PDM Rev. 5.2, SP1 SIMATIC PDM Rev. 5.2 SP1 is supported only for basic configuration and troubleshooting.
Remote operation 7.1 Operating via SIMATIC PDM 1. Click Start/Settings/Control Panel to begin configuration. 2. Double click System, select the Hardware tab, and click the Device Manager button. 3. Open Ports folder and double click the COM Port used by the system to open the Communications Port Properties window. 4. Select the Port Settings tab and double click the Advanced button. 5. If the Use FIFO buffers radio box is selected, click to deselect. 1 ① Deselect Use FIFO buffers radio box 6.
Remote operation 7.1 Operating via SIMATIC PDM 1. Check that you have the most recent EDD, and if necessary update it. [See Updating the Electronic Device Description (EDD) (Page 65) above]. 2. Launch SIMATIC Manager and create a new project for the device. 3. Open the menu Device – Master Reset and click on OK to perform a reset to Factory Defaults. 4. After the reset is complete upload parameters to the PC/PG. 5. Configure the device via the Quick Start wizard. 7.1.
Remote operation 7.1 Operating via SIMATIC PDM Launch SIMATIC PDM, open the menu Device – Wizard - Quick Start, and follow steps 1 to 5. 1 ① Quick Start Step 1 – Identification Note ● The layout of the dialog boxes shown may vary according to the resolution setting for your computer monitor. ● SITRANS PDM limits the TAG field to a maximum of 24 characters. 1.
Remote operation 7.1 Operating via SIMATIC PDM See Application with Stillpipe (Page 60) for a Low Dielectric Liquid application. Step 3 – Vessel Shape The vessel shapes shown are prede‐ fined. For a vessel with obstructions, see Au‐ to False Echo Suppression via SIMAT‐ IC PDM (Page 74). Select the vessel shape, and click on NEXT.
Remote operation 7.1 Operating via SIMATIC PDM Step 4 – Ranges Set the parameters, and click on NEXT. Step 5 – Summary Check parameter settings, and click on BACK to return and revise values, FINISH to save settings offline, or FINISH AND DOWNLOAD to save settings offline and transfer them to the device. The message Quick Start was successful will appear. Click on OK.
Remote operation 7.1 Operating via SIMATIC PDM 7.1.5 Changing parameter settings using SIMATIC PDM Note ● For a complete list of parameters, see Parameter Reference (Page 107). ● Clicking on Cancel during an upload from device to SIMATIC PDM will result in some parameters being updated. Many parameters are accessed via pull-down menus in PDM. See Parameters accessed via pull-down menus (Page 71) for others. 1. Launch SIMATIC PDM, connect to device, and upload data from device. 2.
Remote operation 7.1 Operating via SIMATIC PDM 7.1.6 Parameters accessed via pull-down menus Click on Device or View to open the associated pull-down menus.
Remote operation 7.1 Operating via SIMATIC PDM 7.1.6.2 Echo profile Note ● Double click on each axis to see the Xscale and Data Scale values. ● To zoom in to a section of the profile, left-click and drag a marquee around it. Right click inside the window to zoom out. ● Expand or compress the x and/or y axes: – Left-click on the axis and drag in either direction to reposition the low end of the scale. – Right-click on the axis and drag in either direction to reposition the high end of the scale.
Remote operation 7.1 Operating via SIMATIC PDM Echo profile data logging You can store up to 60 profiles at a selected interval (maximum 60 minutes). Inside Echo Profile Utilities, in the Echo Profile Time Based Storage window: ● Enter the desired interval between stored profiles. ● Enter the maximum number of profiles to be stored (maximum 60). ● Click on Start. A message appears warning of the time delay and warning that all previous saved profiles will be overwritten. Click on OK to proceed.
Remote operation 7.1 Operating via SIMATIC PDM ● Alternatively, enter values for Point and Offset directly into the dialog boxes. ● Click on Transfer to Device. 7.1.6.5 Auto false echo suppression Note ● Ensure material level is below all known obstructions when using Auto False Echo Suppression to learn the echo profile. An empty or almost empty vessel is recommended.
Remote operation 7.1 Operating via SIMATIC PDM 1. Make sure the material level is below all known obstructions. 2. Determine Auto False Echo Suppression Range. Measure the actual distance from the sensor reference point to the material surface using a rope or tape measure. Subtract 0.5 m (20") from this distance, and use the resulting value. 3. Open the menu Device – Echo Profile Utilities and click on the tab Auto False Echo Suppression. 4. Make sure Auto False Echo Suppression Range is On. 5.
Remote operation 7.1 Operating via SIMATIC PDM After Auto False Echo Suppression 3URILOH >G%@ /HYHO GE 4 2 3 5 1 6 'LVWDQFH >P@ ① ② ③ 7.1.6.6 False echo Learned TVT Auto False Echo Suppression Range ④ ⑤ ⑥ Material level Default TVT Echo marker Echo setup Provides quick access to echo selection, filtering, and response rate parameters. Open the menu Device – Echo Profile Utilities and click on Echo Setup.
Remote operation 7.1 Operating via SIMATIC PDM 7.1.6.7 Maintenance You can set schedules and reminders for: ● device maintenance based on its projected lifetime ● sensor maintenance based on its projected lifetime ● service ● calibration To set Device/Sensor Maintenance schedules: 1. Open the menu Device – Maintenance, and click on the Remaining Device/Sensor Lifetime tab. 2.
Remote operation 7.1 Operating via SIMATIC PDM 4. Click Read, to see the effects of your modification. 5. Click Service/Calibration Performed to reset the schedule. 7.1.6.8 Select analog output Allows you to set the mA Output to report Level, Distance, Space, or Volume. See Current Output Function (2.6.1.) for an illustration. If a volume application is selected, mA Output is automatically set to Volume. See Analog Output (Page 230) for more details. 1. Open the menu Device – Select Analog Output. 2.
Remote operation 7.1 Operating via SIMATIC PDM 7.1.6.12 Master reset Factory Defaults Factory Defaults resets all parameters to the default settings with the following exceptions: ● Device Address (5.1.) remains unchanged if the reset command is sent remotely (via AMS, PDM, DTM, FC375) but is reset to 0 if the reset command is sent via LUI. ● Write Protect (6.2.1.) and PIN to Unlock (6.2.2.) ● Learned TVT curve, see Auto False Echo Suppression (2.8.7.1) 1.
Remote operation 7.1 Operating via SIMATIC PDM 7.1.6.16 Trend Open the menu View – Process Variables and click on Trend.
Remote operation 7.
Remote operation 7.1 Operating via SIMATIC PDM 7.1.6.17 Device status Open the menu View – Device Status to view Diagnostics, Device Status, Hardware/ Firmware (HW/FW) Status, and Maintenance status. 1 ① update diagnostics In the Diagnostics window, click on Update diagnostics to update diagnostic information and refresh linked icons. 7.1.6.18 Update Open the menu View – Update to refresh the screen. 7.1.6.
Remote operation 7.2 Operating via AMS Device Manager without it. A maintenance user will be able to open the project without a password but will not have access to security and communication control parameters. 1. Open a project, double-click on the device icon, and in the User window select Specialist. 2. Open the menu Options – Settings and click on the Password tab. 3. Enter a new password and re-enter it in the Confirmation window. Click on OK. 7.
Remote operation 7.2 Operating via AMS Device Manager 7.2.1 Functions in AMS Device Manager Note While the device is in PROGRAM mode the output remains fixed and does not respond to changes in the device. 7.2.1.1 AMS function overview AMS Device Manager monitors the process values, alarms and status signals of the device. It allows you to display, compare, adjust, verify, and simulate process device data.
Remote operation 7.2 Operating via AMS Device Manager 7.2.1.4 Configuring a new device 1. Check our website to make sure you have the most recent EDD. Product page (http://www.siemens.com/LR250) Go to Support > Software Downloads and if necessary download it. Save the files to your computer, and extract the zipped file to an easily accessed location. 2. Launch AMS Device Manager– Add Device Type, browse to the unzipped EDD file and select it. 7.2.1.5 Startup 1. Launch AMS Device Manager.
Remote operation 7.2 Operating via AMS Device Manager 7.2.1.6 Pull-down menu access 1 ① Action menu items A pull-down menu under Actions gives alternative access to several features. Scan Device • Open the menu Actions – Scan Device. • Scan Device uploads parameters from the device (synchronizes parameters). 7.2.1.7 Device configuration 1. Navigate to Configure/Setup > Operation and click to open the dialog window. 2.
Remote operation 7.2 Operating via AMS Device Manager 7.2.1.8 Quick Start Wizard via AMS Device Manager Note ● A reset to Factory Defaults should be performed before running the Quick Start Wizard if device has been used in a previous application. See Master Reset (Page 85). ● The layout of the dialog boxes shown may vary according to the resolution setting for your computer monitor. ● At each step, you can accept the default values without modification and click on the next step to proceed.
Remote operation 7.2 Operating via AMS Device Manager Step 2 – Application 1. Click on Step 2. 2. Select the application type (Level/vessel, Level/stillpipe, Level/bypass pipe, Volume/vessel, Volume/stillpipe, or Volume/bypass pipe) and the material (Liquid or Low dielectric liquid). 3. Click on Apply. Step 3 – Vessel Shape 1. Click on Step 3. 2. Select a predefined vessel shape. To describe a more complex shape see Linearization (Page 88). 3. Click on Apply. Step 4 – Ranges 1. Click on Step 4. 2.
Remote operation 7.2 Operating via AMS Device Manager 1 2 ① ② Breakpoints Maximum volume 1. Click on the Breakpoints tab and enter values for level and volume for up to 32 breakpoints. 2. Navigate to Configure/Setup > Setup > Linearization and click on Characteristic to preview the characteristic curve of the vessel breakpoints. (The conversion will function correctly even if not all breakpoints are entered, but the curve will display correctly only if all breakpoints are entered.) 3.
Remote operation 7.2 Operating via AMS Device Manager Operation Navigate to Configure/Setup > Operation and click on Operation to open the dialog window for access to: General ● Select Analog Output [see Select Analog Output (Page 78)] ● Master Reset [see Master Reset (Page 79)] Simulation/Test ● Self Test [see Self Test (Page 78)] ● Loop Test [see Loop Test (Page 78)] Setup Note For more detailed explanations of the parameters listed below see the pages referenced.
Remote operation 7.2 Operating via AMS Device Manager Sensor Navigate to Configure/Setup > Setup and click on Sensor for access to: General [see Sensor (2.2.)] ● Units ● Operating Mode ● Material ● Damping Filter Calibration [see Calibration (2.3.)] ● Low Calibration Point ● High Calibration Point ● Sensor Offset Rate [see Rate (2.4.)] ● Response Rate ● Fill Rate per Minute ● Empty Rate per Minute Fail-safe [see Fail-safe (2.5.
Remote operation 7.2 Operating via AMS Device Manager Analog Output Scale Navigate to Configure/Setup > Setup and click on Analog Output Scaling for access to: Analog Output Scaling [see Analog Output Scaling (2.6.
Remote operation 7.2 Operating via AMS Device Manager Signal Processing General Navigate to Configure/Setup > Setup > Signal Processing and click on General for access to: Range [see Signal Processing (2.8.)] ● Near Range ● Far Range ● Propagation Factor Echo Select [see Echo Select (2.8.4.)] ● Algorithm ● Position Detect ● Echo Threshold ● CLEF Range Sampling [see Sampling (2.8.5.
Remote operation 7.2 Operating via AMS Device Manager TVT Modify the TVT to screen out false echoes. See Auto False Echo Suppression (2.8.7.1.) (Page 228). Navigate to Configure/Setup > Setup > Signal Processing and click on TVT. Click on one of the two tabs to access the parameters listed: TVT Setup [see TVT setup (2.8.7.)] ● Auto False Echo Suppression ● Auto False Echo Suppression Range ● Hover Level ● Shaper Mode TVT Shaper ● Shaper breakpoints 1 to 40. (Turn TVT Setup/Shaper Mode on to activate.
Remote operation 7.2 Operating via AMS Device Manager Local Display Navigate to Configure/Setup > Setup > Local Display for access to: ● Language ● LCD Fast Mode [see LCD Fast Mode (4.9.)] ● LCD Contrast [see LCD Contrast (4.10.)] 7.2.2 Maintenance and diagnostics Navigate to Maintenance and Diagnostics for access to: Remaining Device Lifetime [see Remaining Device Lifetime (4.2.
Remote operation 7.2 Operating via AMS Device Manager ● Activation of Reminders ● Reminder 1 before Calibration (Required) ● Reminder 2 before Calibration (Demanded) Electronic Temperature ● Electronic Temperature ● Lowest Value ● Highest Value Wear ● Powered Days ● Poweron Resets See Wear via SIMATIC PDM (Page 79) for more detail. 7.2.
Remote operation 7.2 Operating via AMS Device Manager 7.2.5 Device Diagnostics Click on the Device Diagnostics bar at the bottom of the navigation window, for access to: Device Status Hardware/Firmware Status 7.2.5.
Remote operation 7.2 Operating via AMS Device Manager Echo Profile 7.2.5.2 Password protection An AMS Device Manager administrator can configure the user to require a password. The use of passwords is recommended. A password should be assigned to the ’admin’ username immediately after installing AMS Device Manager. Each user is given an AMS Device Manager username and password and required to enter them when they start AMS Device Manager. Access to functions depends on the level of permissions granted.
Remote operation 7.2 Operating via AMS Device Manager The Add User Wizard dialog allows you to: ● select a user type, standard (AMS Device Manager) or Window user.
Remote operation 7.2 Operating via AMS Device Manager 7.2.6 AMS menu structure &RQILJXUH 6HWXS ,GHQWLILFDWLRQ ,GHQWLILFDWLRQ ,GHQWLILFDWLRQ 7DJ 'HVFULSWRU 0HVVDJH 'DWH 'HYLFH 0DQXIDFWXUHU 3URGXFW 1DPH 2UGHU 1XPEHU )LQDO $VVHPEO\ 1XPEHU +DUGZDUH 5HYLVLRQ )LUPZDUH 5HYLVLRQ /RDGHU 5HYLVLRQ ('' YHUVLRQ 'HYLFH 6WDWXV :L]DUG 4XLFN 6WDUW 6WHS 6WHS 6WHS 6WHS 6WHS ,GHQWLILFDWLRQ $SSOLFDWLRQ 9HVVHO 6KDSH 5DQJHV 6XPPDU\ 2SHUDWLRQ 2SHUDWLRQ 7DJ *HQHUDO 6HOHFW $QDORJ 2XWSXW 0DVWHU 5HVHW 6LPXODWL
Remote operation 7.2 Operating via AMS Device Manager &RQILJXUH 6HWXS FRQWLQXHG 6HWXS 6HQVRU FRQWLQXHG 5DWH 5HVSRQVH 5DWH )LOO 5DWH SHU 0LQXWH (PSW\ 5DWH SHU 0LQXWH )DLO VDIH 0DWHULDO /HYHO 7LPHU /HYHO 'HYLFH 6WDWXV $QDORJ 2XWSXW 6FDOLQJ $QDORJ 2XWSXW 6FDOLQJ 7DJ &XUUHQW 2XWSXW )XQFWLRQ P$ 6HWSRLQW P$ 6HWSRLQW 0LQLPXP P$ /LPLW 0D[LPXP P$ /LPLW 'HYLFH 6WDWXV /LQHDUL]DWLRQ 9HVVHO 6KDSH 7DJ 9HVVHO 6KDSH 9HVVHO 6KDSH 9HVVHO 'LPHQVLRQV 0D[LPXP 9ROXPH 'LPHQVLRQ $ 'LPHQVLRQ / 'HYLFH 6WDWXV %UHDNSRL
Remote operation 7.2 Operating via AMS Device Manager &RQILJXUH 6HWXS FRQWLQXHG 6LJQDO 3URFHVVLQJ FRQWLQXHG 797 797 6HWXS 7DJ $XWR )DOVH (FKR 6XSSUHVVLRQ $XWR )DOVH (FKR 6XSSUHVVLRQ 5DQJH +RYHU /HYHO 6KDSHU 0RGH 'HYLFH 6WDWXV 797 6KDSHU 7DJ %UHDNSRLQWV 'HYLFH 6WDWXV 6KDSHU 0RGH 0DQXDO 797 &XUYH 0DQXDO 797 GLDJUDP 7DJ 6KDSHU 0RGH (FKR 3URILOH (FKR 3URILOH 7DJ 'HYLFH 6WDWXV (FKR 3URILOH 3DUDPHWHUV /HYHO 0HDVXUHPHQW 'LVWDQFH 0HDVXUHPHQW &RQILGHQFH 1HDU 5DQJH /RFDO 'LVSOD\ /RFDO 'LVSOD\ 7DJ /DQJXDJH /&'
Remote operation 7.
Remote operation 7.3 Operating via FDT (Field Device Tool) 7.3 Operating via FDT (Field Device Tool) FDT is a standard used in several software packages designed to commission and maintain field devices. Two commercially available FDTs are PACTware and Fieldcare. Functionally FDT is very similar to PDM. See Operating via SIMATIC PDM (Page 63) for more detail. ● To configure a field device via FDT you need the DTM (Device Type Manager) for the device.
Remote operation 7.3 Operating via FDT (Field Device Tool) 7.3.4 Configuring a new device via FDT The full process to configure a field device via FDT is outlined in an Application Guide which can be downloaded from our website under Support > Application Guides. Product page (http://www.siemens.
Remote operation 7.
Parameter reference 8 Note ● Parameter names and menu structure are almost identical for SIMATIC PDM and the local user interface (LUI). ● Default settings in the parameter tables are indicated with an asterisk (*) unless explicitly stated. ● Mode toggles between PROGRAM and Measurement Modes. ● For Quick Access to parameters via the handheld programmer, press Home , then enter the menu number, for example: 2.2.1. ● In Navigation mode, ARROW keys ( ) navigate the menu in the direction of the arrow.
Parameter reference Setup (2.) Note Default settings in the parameter tables are indicated with an asterisk (*) unless explicitly stated. Device (2.1.) Hardware Revision (2.1.1.) Corresponds to the electronics hardware of the SITRANS LR250. Firmware Revision (2.1.2.) Corresponds to the firmware that is embedded in the SITRANS LR250. Loader Revision (2.1.3.) Corresponds to the software used to update the SITRANS LR250. Order Option (2.1.4.) Read only.
Parameter reference Sensor Mode (2.2.2.) 5 3 4 2 1 8 7 6 ① Sensor value ⑤ ② ③ ④ Space ⑥ ⑦ ⑧ Distance High Calibration Point Sensor reference point [see Dimension draw‐ ings] (Page 175) Low Calibration Point Far range Level Mode Description Reference point NO SERVICE Measurement and associated loop current are not updated, and the device defaults to Failsafe mode a).
Parameter reference Damping Filter (2.2.4.) The time constant for the damping filter. The damping filter smooths out the response to a sudden change in level. This is an exponential filter and the engineering unit is always in seconds [see Damping (Page 230) for more detail]. Values Range: 0 to 100.000 s Default: 10.000 s Calibration (2.3.) Note We recommend using the Quick Start wizard to configure the device. Low Calibration Pt. (2.3.1.) Distance from sensor reference point1) to Low Calibration Point.
Parameter reference Sensor Offset (2.3.3.) A constant offset that can be added to or subtracted from the sensor value1) to compensate for a shifted sensor reference point. (For example, when adding a thicker gasket or reducing the standoff/nozzle height.) The units are defined in Units (2.2.1.). Values Range: -100 to 100 m. Default: 0.00 m Related parameters Units (2.2.1.
Parameter reference Sensor value is the value produced by the echo processing which represents the distance from sensor reference point to the target. See Sensor Mode (2.2.2.) for an illustration. Empty Rate per Minute (2.4.3.) Defines the maximum rate at which the reported sensor value is allowed to increase. Adjusts the SITRANS LR250 response to increases in the actual material level. Empty Rate is automatically updated whenever Response Rate is altered. Options Range: 0 to 99999 m / min.
Parameter reference NAMUR NE 43-COMPLIANT DEVICE1) Options * 1) HI 20.5 mA (max. mA Limit) LO 3.8 mA (min. mA Limit) HOLD Last valid reading VALUE User-selected value [defined in Fail-Safe mA Value (2.5.3.): default 3.58 mA] Orderable option LOE Timer (2.5.2.) Note When a Loss of Echo occurs Material Level (2.5.1.) determines the material level to be reported when the Fail-safe timer expires. See Loss of Echo (LOE) (Page 232) for more detail.
Parameter reference Current Output Function (2.6.1.) Note ● Level, space, and distance, have different reference points. ● Use caution when changing Current Output Function while the device is connected to a HART network. Current Output Function controls the primary value and the loop current for the device. ● Current Output Function also affects the secondary, tertiary and quaternary variables in a HART network. Can be set to either Level, Space, Distance, or Volume.
Parameter reference 4 mA Setpoint (2.6.2.) Sets the process level corresponding to the 4 mA value. 4 mA always defaults to 0, and Current Output Function (2.6.1.) determines whether this is a Level, Space, Distance, or Volume measurement. (See Current Output Function (2.6.1.) for an illustration.) Values Range: –999999 to +999999 (limits vary with current function and units) Default: 0.00 m (set to value corresponding to 0% as defined by Current Output Function) Related Parameters Units (2.2.1.
Parameter reference mA Output Value (2.6.6.) Allows you to use a simulated value to test the functioning of the loop. You can enter 4 mA, 20 mA, or any other user-defined value within the range. Values Range: 3.56 mA to 22.6 mA Read Only unless Current Output Function (2.6.1.) is set to Manual. Related parameter Current Output Function (2.6.1.) 1. First set Current Output Function (2.6.1.) to Manual. 2. Enter the desired mA value in mA Output Value. 3.
Parameter reference Vessel Shape LCD DISPLAY/ Description Also required LINEAR/ Upright, linear (flat bottom) Maximum volume CONICAL BOT/ Conical or pyramidal bot‐ tom Maximum volume, di‐ mension A $ PARABOLIC BOT/Parabol‐ Maximum volume, di‐ ic bottom mension A $ HALF SPHERE BOT/ Half-sphere bottom Maximum volume, di‐ mension A FLAT SLOPED BOT/ Flat sloped bottom Maximum volume, di‐ mension A $ $ SITRANS LR250 with mA/HART Operating Instructions, 10/2019, A5E32220602-AG 117
Parameter reference Vessel Shape LCD DISPLAY/ Description Also required PARABOLIC ENDS/ Parabolic end horizontal cylinder $ Maximum volume, di‐ mension A, dimension L / LINEAR TABLE a) / Lineari‐ Maximum volume, ta‐ zation table (level/volume bles 1-32 level and vol‐ breakpoints) ume breakpoints a) Linearization Table must be selected in order for level/volume values [see Table 1-8 (2.7.2.)] to be transferred. Maximum Volume (2.7.1.2.) The maximum volume of the vessel.
Parameter reference Vessel Dimension L (2.7.1.4.) Length of the cylindrical section of a horizontal parabolic end vessel. See Vessel Shape (2.7.1.1.) for an illustration. Values Range: 0.0 to 99.99 m Default: 0.0 Related Parameters Units (2.2.1.) Vessel Shape (2.7.1.1.) Table 1-8 (2.7.2.) Note Linearization Table must be selected in Vessel Shape (2.7.1.1.) in order for level/volume values to be transferred.
Parameter reference Level 1 (2.7.2.1.) 1. Press RIGHT arrow to open Edit mode. 2. Enter level value and press RIGHT arrow to except it. 3. Press DOWN arrow to move to corresponding volume breakpoint. Volume 1 (2.7.2.2.) 1. Press RIGHT arrow to open Edit mode. 2. Enter volume value and press RIGHT arrow to accept it. 3. Press DOWN arrow to move to next level breakpoint.
Parameter reference Signal Processing (2.8.) Note Default settings in the parameter tables are indicated with an asterisk (*) unless explicitly stated. Near Range (2.8.1.) The range in front of the device (measured from the sensor reference point) within which any echoes will be ignored. This is sometimes referred to as blanking or a dead zone. The factory setting is 50 mm (2") past the end of the antenna, and the default is dependent on the antenna type and process connection.
Parameter reference Propagation Factor 0.9844 0.988 0.9935 0.9965 CLEF Range (2.8.4.4.) settings Low calibra‐ tion point 700 mm (2.29 ft)b) Low calibra‐ tion point 700 mm (2.29 ft)b) Low calibration point -1000 mm (3.28 ft)b) Low calibration point -1000 mm (3.28 ft)b) a) Since pipe dimensions may vary slightly, the propagation factor may also vary.
Parameter reference Position Detect (2.8.4.2.) Defines where on the echo the distance measurement is determined. Options Center * Hybrid (Center and CLEF) CLEF (Constrained Leading Edge Fit) Related parameters CLEF Range (2.8.4.4.) If the vessel bottom is being reported as the level instead of the actual material level (at low level conditions), or if the dielectric constant of the liquid to be monitored is less than 3, we recommend setting Position Detect to Hybrid and CLEF Range (2.8.4.4.) to 0.
Parameter reference 2 1 ① ② ③ 3 CLEF Range Sensor reference point Low Calibration Point (process empty level ) In applications with low dK materials we recommend setting CLEF Range to 0.5 m (1.64 ft) and Position Detect (2.8.4.2.) to Hybrid. Sampling (2.8.5.) Provides a method of checking the reliability of a new echo before accepting it as the valid reading, based on numbers of samples above or below the currently selected echo. Echo Lock (2.8.5.1.
Parameter reference Sampling Down (2.8.5.3.) Specifies the number of consecutive echoes that must appear below the echo currently selected, before the measurement is accepted as valid. Values Range: 1 to 50 Default: 2 Echo Quality (2.8.6.) Confidence (2.8.6.1.) Indicates echo reliability: higher values represent better echo quality. The display shows the echo confidence of the last measurement. Echo Threshold (2.8.4.3.) defines the minimum criterion for echo confidence.
Parameter reference Auto False Echo Suppression (2.8.7.1.) Used together with Auto False Echo Suppression Range (2.8.7.2.) to screen out false echoes in a vessel with known obstructions. A ’learned TVT’ (time varying threshold) replaces the default TVT over a specified range. See Auto False Echo Suppression (Page 228) for a more detailed explanation. Note ● Make sure material level is below all known obstructions when Auto False Echo Suppression is used to learn the echo profile.
Parameter reference 4. Go to Auto False Echo Suppression (2.8.7.1.) and press RIGHT arrow to open Edit Mode. 5. Select Learn. The device will automatically revert to On (Use Learned TVT) after a few seconds.
Parameter reference Specifies the range within which Learned TVT is used [see Auto False Echo Suppression (2.8.7.1.) for more detail]. Values Range: 0.00 to 20.00 m Default: 1.00 m Related parameters Units (2.2.1.) 1. Calculate range according to Auto False Echo Suppression (2.8.7.1.) steps 1 and 2. 2. Press RIGHT arrow to open Edit mode. 3. Enter the new value and press RIGHT arrow to accept it. 4. Set Auto False Echo Suppression (2.8.7.1.). Hover Level (2.8.7.3.
Parameter reference To use TVT shaper via SIMATIC PDM: 1. Go to Level Meter > Setup > Signal Processing > TVT setup > Shaper Mode and select On. 2. Open the menu Device – Echo Profile Utilities and click on TVT Shaper. For more detail see TVT shaper via SIMATIC PDM (Page 73). To use TVT shaper via LUI (local user interface): 1. Go to Shaper Mode (2.8.7.4.) and select On. 2. Go to Breakpoint 1-9 (2.8.8.1.). 3. Open Shaper 1 and enter the TVT Offset value (between –50 and 50). 4.
Parameter reference To access measured values via SIMATIC PDM: Open the menu View – Process Variables. Level Measurement (2.8.9.1.) The value for level. Space Measurement (2.8.9.2.) The value for space. Distance Measurement (2.8.9.3.) The value for distance. Volume Measurement (2.8.9.4.) The value for volume. Diagnostics (3.) Echo Profile (3.1.) Allows you to request the current echo profile either locally via the handheld programmer, or remotely via SIMATIC PDM, or AMS Device Manager.
Parameter reference For access via SIMATIC PDM open the menu View – Process Variables and check the field Electronics Temperature. Highest Value (3.2.2.) Read only. Displays (in degrees C) the maximum temperature recorded by the internal electronics. The high and low values are maintained over a power cycle. Via SIMATIC PDM navigate to Maintenance and Diagnostics > Electronics Temperature. Lowest Value (3.2.3.) Read only.
Parameter reference Remaining Device Lifetime (4.2.) Note ● Default settings in the parameter tables are indicated with an asterisk (*) unless explicitly stated. ● Four sets of parameters allow you to monitor the Device/Sensor Lifetimes and set up Maintenance/Service schedules, based on operating hours instead of a calendar-based schedule. See also Remaining Sensor Lifetime (4.3.), Service Schedule (4.4.), and Calibration Schedule (4.5.).
Parameter reference Lifetime (Expected) (4.2.1.) Note The device always operates in years. Changing the units affects only the parameter view of the Service Interval parameters in SIMATIC PDM. User-configurable recommended time between product inspections. Values Units a): hours, days, years Range: 0 to 20 years Default: 10 years a) Units are selectable only via SIMATIC PDM. Time in Operation (4.2.2.) Read only. The amount of time the device has been operating. Remaining Lifetime (4.2.3.) Read only.
Parameter reference Activation of Reminders (4.2.4.) Note To modify this parameter via SIMATIC PDM it must be accessed via the pull-down menu Device – Maintenance. Allows you to enable a maintenance reminder. Values * Timer OFF ON - no reminders checked ON - Reminder 1 (Maintenance Required) checked ON - Reminders 1 and 2 checked ON - Reminder 2 (Maintenance Demanded) checked 1. First set the values in Reminder 1 before Lifetime (Required) (4.2.5.)/ Reminder 2 before Lifetime (Demanded) (4.2.6.). 2.
Parameter reference Acknowledge Status (4.2.8.) Indicates which level of maintenance reminder has been acknowledged. In SIMATIC PDM, open the menu View – Device Status, click on the Maintenance tab, and check the Device Lifetime Status window. Acknowledge (4.2.9.) Acknowledges the current maintenance reminder. To acknowledge a reminder via SIMATIC PDM: 1. Open the menu View – Device Status and click on the Maintenance tab. 2. In the Device Lifetime section, click on Acknowledge Warnings.
Parameter reference Time Units Options a) Hours; days; years Default: years a) Selectable only via SIMATIC PDM. Lifetime (Expected) (4.3.1.) Note The device always operates in years. Changing the units affects only the parameter view of Remaining Sensor Life parameters in SIMATIC PDM. Allows you to override the factory default. Values Units a): hours, days, years Range: 0 to 20 years Default: 10.00 years a) Units are selectable only via SIMATIC PDM. Time in Operation (4.3.2.
Parameter reference To reset to zero: ● In SIMATIC PDM, open the menu Device – Maintenance, click on the Remaining Sensor Lifetime tab, and click on Sensor Replaced to restart the timer and clear any fault messages. ● Via the handheld programmer, manually reset Time in Operation (4.3.2.) to zero. Remaining Lifetime (4.3.3.) Read only. Lifetime (Expected) (4.3.1.) less Time in Operation (4.3.2.). Activation of Reminders (4.3.4.
Parameter reference 1. Modify values as required. 2. Set Activation of Reminders (4.3.4.) to the desired option. Maintenance Status (4.3.7.) Indicates which level of maintenance reminder is active. In SIMATIC PDM, open the menu View – Device Status, click on the Maintenance tab, and check the Sensor Lifetime Status window. Acknowledge Status (4.3.8.) Indicates which level of maintenance reminder has been acknowledged.
Parameter reference optimal use, we recommend that you use SIMATIC PCS7 Asset Management Software in conjunction with SIMATIC PDM. To access these parameters via SIMATIC PDM: ● Open the menu Device – Maintenance and select the Service Schedule tab. ● After modifying values/units as required, click on Write to accept the change, and Read to view the effect of the change. ● Click on Service Performed to restart the timer and clear any fault messages.
Parameter reference User-configurable recommended time between product inspections. Values Units a): hours, days, years Range: 0 to 20 years Default: 1.0 year a) Units are selectable only via SIMATIC PDM. Time Since Last Service (4.4.2.) Time elapsed since last service. Can be reset to zero after performing a service.
Parameter reference 1. Modify values as required. 2. Set Activation of Reminders (4.4.4.) to the desired option. Reminder 2 before Service (Demanded) (4.4.6.) If Time Until Next Service (4.4.3.) is equal to or less than this value, the device generates a Maintenance Required reminder. Values Range: 0 to Service Interval (4.4.1.) Default: 0.019 years 1. Modify values as required 2. Set Activation of Reminders (4.4.4.) to the desired option. Maintenance Status (4.4.7.
Parameter reference Calibration Schedule (4.5.) Note ● Default settings in the parameter tables are indicated with an asterisk (*) unless explicitly stated. ● Four sets of parameters allow you to monitor the Device/Sensor Lifetimes and set up Maintenance/Service schedules, based on operating hours instead of a calendar-based schedule. See also Remaining Device Lifetime (4.2.), Remaining Sensor Lifetime (4.3.), and Service Schedule (4.4.).
Parameter reference Calibration Interval (4.5.1.) Note The device always operates in years. Changing the units affects only the parameter view of Remaining Sensor Life parameters in SIMATIC PDM. User-configurable recommended time between product calibrations. Values Units a): hours, days, years Range: 0 to 20 years Default: 1.0 year a) Units are selectable only via SIMATIC PDM. Time Since Last Calibration (4.5.2.) Time elapsed since last calibration.
Parameter reference Activation of Reminders (4.5.4.) Note To modify this parameter via SIMATIC PDM it must be accessed via the pull-down menu Device – Maintenance. Allows you to enable a maintenance reminder. Options Timer OFF ON - no reminders checked ON - Reminder 1 (Maintenance Required) checked * ON - Reminders 1 and 2 checked ON—Reminder 2 (Maintenance Demanded) checked 1. First set the values in Reminder 1 before Calibration (Required) (4.5.5.)/Reminder 2 before Calibration (Demanded) (4.5.6.). 2.
Parameter reference Acknowledge Status (4.5.8.) Indicates which level of maintenance reminder has been acknowledged. In SIMATIC PDM, open the menu View – Device Status, click on the Maintenance tab and check the Calibration Schedule Status window. Acknowledge (4.5.9.) Acknowledges the current maintenance reminder. To acknowledge a reminder via SIMATIC PDM: 1. Open the menu View – Device Status and click on the Maintenance tab. 2. In the Service Schedule Status section, click on Acknowledge Warnings.
Parameter reference LCD Contrast (4.10.) The factory setting is for optimum visibility at room temperature and in average light conditions. Extremes of temperature will lessen the contrast. Values Range: 0 (low contrast) to 20 (high contrast). Default: 10 Adjust the value to improve visibility at room temperature and in average light conditions. Change the value in small steps to ensure you can continue to read the display. Secondary Value (4.11.
Parameter reference Device Address (5.1.) Sets the device address or poll ID on a HART network. Any address other than 0 will cause the output current to be a fixed value, and the current will not indicate the reading. Values Range: 0 to 15 Default: 0 To set Device Address via SIMATIC PDM: ● Open the project in Process Device Network View then right-click on the device. ● Go to Object Properties > Connection to access the field Short Address. Security (6.
Parameter reference Write Protect (6.2.1.) Note This lock affects only the handheld programmer. A remote master can change configuration if Access Control (6.1.1.) is set to allow this. Prevents any changes to parameters via the handheld programmer. Options Range: 0 to 9999 * Unlock value [stored in PIN to Unlock (6.2.2.)] Lock Off Any other value Lock On ● To turn Lock On, key in any value other than the Unlock Value stored in PIN to Unlock (6.2.2.).
Parameter reference 8.1 Alphabetical parameter list 8.1 Alphabetical parameter list Note For a detailed list of parameters see Parameter Reference (Page 107). Maintenance Parameters are not listed below. See Remaining Device Lifetime (4.2.), Remaining Sensor Lifetime (4.3.), Service Schedule (4.4.) and Calibration Schedule (4.5.) for those parameters. 20 mA Setpoint (2.6.3.) 4 mA Setpoint (2.6.2.) Access Control (6.1.1.) Algorithm (2.8.4.1.) Analog Output Scaling (2.6.) Auto False Echo Suppression (2.8.7.
Parameter reference 8.1 Alphabetical parameter list Electronics Temperature (3.2.) Empty Rate per Minute (2.4.3.) Fail-safe (2.5.) Fail-Safe mA Value (2.5.3.) Far Range (2.8.2.) Fill Rate per Minute (2.4.2.) Firmware Revision (2.1.2.) Hardware Revision (2.1.1.) High Calibration Pt. (2.3.2.) Highest value (3.2.2.) Hover Level (2.8.7.3.) Language (7.) LCD Contrast (4.10.) LCD Fast Mode (4.9.) Level 1 (2.7.2.1.) Level Measurement (2.8.9.1.) Linearization (2.7.) Loader Revision (2.1.3.) Local Access (6.2.
Parameter reference 8.1 Alphabetical parameter list Order Option (2.1.4.) PIN to Unlock (6.2.2.) Position Detect (2.8.4.2.) Power-on Resets (4.8.) Powered Hours (4.7.) Propagation Factor (2.8.3.) Quick Start (1.) Quick Start Wizard (1.1.) Rate (2.4.) Remote Access (6.1.) Response Rate (2.4.1.) Sampling (2.8.5.) Sampling Down (2.8.5.3.) Sampling Up (2.8.5.2.) Shaper Mode (2.8.7.4.) Signal Processing (2.8.) Secondary Value (4.11.) Security (6.) Sensor (2.2.) Sensor Mode (2.2.2.) Sensor Offset (2.3.3.
Parameter reference 8.1 Alphabetical parameter list Volume 1 (2.7.2.2.) Volume Measurement (2.8.9.4.) Write Protect (6.2.1.
Service and maintenance 9.1 9 Basic safety information WARNING Impermissible repair of the device ● Repair must be carried out by Siemens authorized personnel only. CAUTION Releasing button lock Improper modification of parameters could influence process safety. ● Make sure that only authorized personnel may cancel the button locking of devices for safety-related applications. 9.2 Cleaning The radar device requires no cleaning under normal operating conditions.
Service and maintenance 9.3 Maintenance and repair work WARNING Electrostatic charge Danger of explosion in hazardous areas if electrostatic charges develop, for example, when cleaning plastic surfaces with a dry cloth. ● Prevent electrostatic charging in hazardous areas. 9.3 Maintenance and repair work The device is maintenance-free. However, a periodic inspection according to pertinent directives and regulations must be carried out.
Service and maintenance 9.3 Maintenance and repair work 9.3.1 Unit repair and excluded liability All changes and repairs must be done by qualified personnel, and applicable safety regulations must be followed. Please note the following: ● The user is responsible for all changes and repairs made to the device. ● All new components must be provided by Siemens. ● Restrict repair to faulty components only. ● Do not re-use faulty components. 9.3.
Service and maintenance 9.
Service and maintenance 9.5 Return procedure 9.4 Disposal Devices described in this manual should be recycled. They may not be disposed of in the municipal waste disposal services according to the Di‐ rective 2012/19/EC on waste electronic and electrical equipment (WEEE). Devices can be returned to the supplier within the EC, or to a locally ap‐ proved disposal service for eco-friendly recycling. Observe the specific regulations valid in your country.
Service and maintenance 9.
10 Diagnosing and troubleshooting 10.1 Communication troubleshooting 1. Check the following: – There is power at the device. – The LCD shows the relevant data. – The device can be programmed using the handheld programmer. – If any fault codes are being displayed see General Fault Codes (Page 160) for a detailed list. 2. Verify that the wiring connections are correct. 3. See the table below for specific symptoms. Symptom Corrective action The device cannot be programmed via the handheld programmer.
Diagnosing and troubleshooting 10.
Diagnosing and troubleshooting 10.3 General fault codes Code/ Icon S: 0 * Meaning Corrective Action The device was unable to get a measurement within the Fail-safe LOE Timer period. Possible causes: faulty installation, antenna material buildup, foaming/other adverse process condi‐ tions, invalid configuration range. ● Ensure installation details are correct. ● Ensure no antenna material buildup. Clean if necessary. ● Adjust process conditions to minimize foam or other adverse conditions.
Diagnosing and troubleshooting 10.3 General fault codes Code/ Icon Meaning Corrective Action S: 18 Calibration interval as defined in Maintenance Demanded Limit has expired. Perform calibration. S: 28 * Internal device failure caused by a RAM memo‐ Repair required: contact your local Siemens representative. ry error. S: 29 * EEPROM damaged. Repair required: contact your local Siemens representative S: 31 * Flash error.
Diagnosing and troubleshooting 10.4 Operation troubleshooting Code/ Icon Meaning Corrective Action S: 49 * EEPROM corrupt. Repair required: contact your local Siemens representative. S: 50 * EEPROM corrupt. Repair required: contact your local Siemens representative. S: 51 * EEPROM corrupt. Repair required: contact your local Siemens representative. Fail-safe is activated. Possible causes: For 3: 1. hardware failure ● Correct configuration; ensure installation is correct; 2.
Diagnosing and troubleshooting 10.4 Operation troubleshooting Symptom Cause Action Display shows antenna malfunction: ● check Current Internal Temperature (3.2.1.) ● temperature too high ● use foam deflector or stillpipe ● physical damage ● relocate ● excessive foam ● use a defoamer 6 /2( Reading does not change, but the level does ● multiple echoes ● set Algorithm (2.8.4.1.
11 Technical data Note Device specifications Siemens makes every attempt to ensure the accuracy of these specifications but reserves the right to change them at any time. 11.1 Power General Purpose Intrinsically Safe Non-Sparking Non-incendive (FM/CSA US/Canada only) Nominal 24 V DC at 550 Ohm Flameproof Increased Safety Explosion-proof (FM/CSA US/Canada only) Nominal 24 V DC at 250 Ohm ● Maximum 30 V DC ● 4 to 20 mA ● Max. startup current see Startup Behaviour (Page 246). 11.
Technical data 11.2 Performance Maximum measurement range4) 1.5" antenna 10 m (32.8 ft)5) 2" threaded PVDF antenna 2"/DN50/50A Flanged encapsulated antenna (FEA) 2" ISO 2852, DN50 DIN 11864-1/2/3, DN50 DIN11851, Tuchenhagen Types F and N Hygienic encapsula‐ ted antenna (HEA) all other versions Minimum detectable distance 20 m (65.
Technical data 11.3 Interface 11.3 Analog output Communication: HART1) Configuration Interface Signal range 4 to 20 mA (± 0.02 mA accuracy) upper limit 20 to 23 mA adjustable Fail signal 3.6 mA to 23 mA [For more details, see Fail-safe Mode] (Page 232) Load 230 to 600 Ω, 230 to 500 Ω when connecting a coupling module Max. line length multi-wire: ≤ 1500 m (4921 ft) Protocol HART, Version 5.
Technical data 11.4 Mechanical 11.4 Transmitter housing Mechanical Stainless steel horn version Process connec‐ tion 2, 3, 4" (ASME B16.5 raised face 150 lb, 300 lb) DN50, DN80, DN100, DN150 (EN 1092-1 TYPE B1 raised face PN10/16, PN25/40), 50A, 80A, 100A (10K JIS B 2220 flat faced) Threaded 1.5" NPT (ASME B1.20.1), R (BSPT, EN 10226‑1), G (BSPP, EN ISO 228‑1), 2" NPT (ASME B1.20.1), R (BSPT, EN 10226‑1), G (BSPP, EN ISO 228‑1), 3" NPT (ASME B1.20.
Technical data 11.4 Mechanical Weight (excluding extensions) 1.5" threaded connection with 1.5" horn antenna approximately 5.1kg (11.2 lb) 2" threaded connection with 2" horn antenna approximately 5.5 kg (12.1 lb) 3" threaded connection with 3" horn antenna approximately 7.0 kg (15.4 lb) Aluminum horn, with polypropylene lens antenna version with 3" flange approximately 3.4 kg (7.5 lb) 2" threaded PVDF antenna approximately 3.3 kg (7.
Technical data 11.6 Process a) For use with 1.5" (40 mm) horn antennas only. 11.5 Environmental Note ● For the specific configuration you are about to use or install, check transmitter nameplate and see Approvals (Page 171). ● Use appropriate conduit seals to maintain IP or NEMA rating. Location indoor/ outdoor Altitude 5000 m (16,404 ft) max.
Technical data 11.
Technical data 11.7 Approvals Application type Hazardous LR250 version Intrinsically safe (Page 40) Valid for: Europe IECEx SIR 05.0031X, Ex ia IIC T4 Ga Ex ia ta IIIC T100 °C Da International FM/CSA Class I, Div. 1, Groups A, B, C, D Class II, Div. 1, Groups E, F, G Class III T4 US/Canada INMETRO: DNV 12.
Technical data 11.8 Programmer (infrared keypad) Application type LR250 version Approval rating Valid for: ABNT NBR IEC 60079-18:2010, ABNT NBR IEC 60079-26:2008, ABNT NBR IEC 60079-31:2011 Flameproof (Page 42)/ NEPSI Increased safety (Page 42) Ex d ia mb IIC T4 Ga/Gb / Ex e ia mb IIC T4 Ga/Gb Ex iaD 20 T90 IP67 DIP A20 TA 90 ºC Explosion proof (Page 43) FM/CSA Class I, Div. 1, Groups A, B, C, D Class II, Div.
Technical data 11.8 Programmer (infrared keypad) Siemens Milltronics Infrared IS (Intrinsically Safe) Handheld Programmer for hazardous and all other locations (battery is non-replaceable). Approvals Ambient temperature Interface Power Weight Color Part number 174 CE FM/CSA Class I, II, III, Div. 1, Gr.
12 Dimension drawings 12.1 Universal polymeric flange r $ $ %&' ① ② ③ ④ Bolt hole circle max. diameter Bolt hole circle min. diameter Bolt hole Flange O.D.
Dimension drawings 12.1 Universal polymeric flange Nominal pipe size Flange O.D. Thickness Bolt hole cir‐ cle max. ∅ Bolt hole cir‐ cle min. Ø Bolt hole ra‐ dius No. of slot‐ ted holes 3 (80) 7.87 (200) 0.7 (17.5) 6.30 (160) 5.91 (150) 0.37 (9.5) 8 4 (100) 9.00 (229) 0.7 (17.5) 7.52 (191) 6.89 (175) 0.37 (9.5) 8 6 (150) 11.22 (285) 0.7 (17.5) 9.53 (242) 9.45 (140) 0.45 (11.
Dimension drawings 12.2 Aluminum horn antenna 12.
Dimension drawings 12.3 Mounting bracket 12.3 Mounting bracket º [ ① ② ③ ④ 178 Stainless steel socket head screw (4) M5 x 0.8 x 14 screws, washers, lock washers Mounting bracket (2) M5 x 0.8 x 16 screws, washers, lock washers (2) M5 x 0.
Dimension drawings 12.3 Mounting bracket Dimensions in mm (inch) Note Torque the bracket screws to 5 Nm (3.7ft/lbs) to hold the aiming adjustment of the antenna.
Dimension drawings 12.4 Aluminum horn with polypropylene lens antenna and polymeric flange 12.
Dimension drawings 12.5 Stainless steel threaded horn antenna 12.5 Stainless steel threaded horn antenna Note ● Process temperature and pressure capabilities are dependent upon information on the process connection tag. Reference drawing listed on the tag is available for download from our website under Support/Installation drawings/Level Measurement/Continuous - Radar/ LR250: Product page (http://www.siemens.
Dimension drawings 12.5 Stainless steel threaded horn antenna 1 2 7 8 11 9 3 4 5 10 6 ① ② ③ ④ ⑤ ⑥ ½" NPT cable entry, or M20 cable gland threaded cover 2" horn 3" horn 4" horn ⑦ ⑧ ⑨ ⑩ ⑪ enclosure/electronics retaining collar process connection tag horn sensor reference point horn O.D. Dimensions in mm (inch) Threaded horn dimensions Antenna Type Antenna O.D.
Dimension drawings 12.5 Stainless steel threaded horn antenna a) Height from bottom of horn to sensor reference point as shown: see dimension drawing. -3dB in the direction of the polarization axis. For an illustration, see Polarization reference point (Page 29).
Dimension drawings 12.6 Stainless steel threaded horn antenna with extension 12.6 Stainless steel threaded horn antenna with extension ① ② ③ ④ ⑤ ⑥ ½" NPT cable entry, or M20 cable gland threaded cover 2" horn 3" horn 4" horn ⑦ ⑧ ⑨ ⑩ ⑪ enclosure/electronics retaining collar process connection tag horn sensor reference point horn O.D.
Dimension drawings 12.6 Stainless steel threaded horn antenna with extension Threaded horn with extension dimensions Antenna Type Antenna O.D., mm (inch) 1-1/2" threaded con‐ nection 2" threaded con‐ nection Beam An‐ b) 3" threaded con‐ gle (°) nection Measurement range, m (ft) 1.5" 39.8 (1.57) 235 (9.25) N/A N/A 19 10 (32.8) 2" 47.8 (1.88) N/A 266 (10.47) 280 (11.02) 15 20 (65.6) 3" 74.8 (2.94) N/A 299 (11.77) 313 (12.32) 10 20 (65.6) 4" 94.8 (3.73) N/A 354 (13.94) 368 (14.
Dimension drawings 12.7 Stainless steel flanged horn antenna 12.7 Stainless steel flanged horn antenna 1 2 5 7 8 6 10 3 9 4 ① ② ③ ④ ⑤ ½" NPT cable entry, or M20 cable gland threaded cover horn horn O.D.
Dimension drawings 12.7 Stainless steel flanged horn antenna -3dB in the direction of the polarization axis (see Polarization reference point (Page 29) for an illustration). b) Optional alloy N06022/2.4602 (Hastelloy® C-22 or equivalent). See Raised-Face Flange Dimensions (Page 213). c) Note Heights to sensor reference point are for stainless steel flanges. For optional alloy N06022/2.4602 (Hastelloy© C-22 or equivalent) see Flanged Horn dimensions above.
Dimension drawings 12.8 Stainless steel flanged horn antenna with extension 12.8 Stainless steel flanged horn antenna with extension 1 2 5 7 8 6 10 9 3 4 ① ② ③ ④ ⑤ ½" NPT cable entry, or M20 cable gland threaded cover horn horn O.D.
Dimension drawings 12.8 Stainless steel flanged horn antenna with extension Flanged horn with extension dimensions Nominal horn size, mm (inch) Horn O.D., mm (inch) 50 (2) 80 (3) 100 (4) Height to sensor reference point,, mm (inch) Beam angle (°)b) Measurement range, m (ft) a) Stainless steel flange: raised or flat-face Optional alloy flange c) 47.8 (1.88) 235.3 (9.26) 238.3 (9.38) 15 74.8 (2.94) 268.3 (10.56) 271.3 (10.68) 10 94.8 (3.73) 323.3 (12.73) 326.3 (12.85) 8 20 (65.
Dimension drawings 12.9 PTFE Flanged encapsulated antenna (2"/DN50/50A sizes only) 12.
Dimension drawings 12.9 PTFE Flanged encapsulated antenna (2"/DN50/50A sizes only) Flange size Flange class Flange O.D. [mm (inch)] Antenna aperture size [mm (inch)] Beam angle (°)1) Measurement range [m (ft)] 50 (1.97) 12.8 10 (32.8)2) 2" 150 LB 152 (5.98) DN50 PN10/16 165 (6.50) 50A 10K 155 (6.10) 1) -3 dB in the direction of the polarization axis. 2) 20m if installed in stillpipe See Raised-Face Flange per EN 1092-1, (Page 215)and Polarization reference point (Page 29).
Dimension drawings 12.10 PTFE Flanged encapsulated antenna (3"/DN80/80A sizes and larger) 12.
Dimension drawings 12.10 PTFE Flanged encapsulated antenna (3"/DN80/80A sizes and larger) Flange size Flange class Flange O.D. [mm (inch)] Antenna aperture size [mm (inch)] Beam angle (°)1) Measurement range [m (ft)] 75 (2.95) 9.6 20 (65.6) 75 (2.95) 9.6 20 (65.6) 75 (2.95) 9.6 20 (65.6) 3" 150 LB 190 (7.48) DN80 PN10/16 200 (7.87) 80A 10K 185 (7.28) 4" 150 LB 230 (9.06) DN100 PN10/16 220 (8.66) 100A 10K 210 (8.27) 6" 150 LB 280 (11.02) DN150 PN10/16 285 (11.
Dimension drawings 12.11 Hygienic encapsulated antenna (2" ISO 2852 sanitary clamp) 12.11 Hygienic encapsulated antenna (2" ISO 2852 sanitary clamp) 137 FDEOH HQWU\ RU DOWHUQDWLYHO\ 0 FDEOH JODQG 7KUHDGHG FRYHU (QFORVXUH 5HWDLQLQJ FROODU 3URFHVV FRQQHFWLRQ 6HQVRU UHIHUHQFH SRLQW Dimensions in mm (inch) 194 SITRANS LR250 with mA/HART Operating Instructions, 10/2019, A5E3222060
Dimension drawings 12.12 Hygienic encapsulated antenna (3" ISO 2852 sanitary clamp) 12.12 Hygienic encapsulated antenna (3" ISO 2852 sanitary clamp) 137 FDEOH HQWU\ RU DOWHUQDWLYHO\ 0 FDEOH JODQG 7KUHDGHG FRYHU (QFORVXUH 5HWDLQLQJ FROODU 3URFHVV FRQQHFWLRQ 6HQVRU UHIHUHQFH SRLQW Dimensions in mm (inch) SITRANS LR250 with mA/HART Operating Instructions, 10/2019, A5E32220602-
Dimension drawings 12.13 Hygienic encapsulated antenna (4" ISO 2852 sanitary clamp) 12.13 Hygienic encapsulated antenna (4" ISO 2852 sanitary clamp) 137 FDEOH HQWU\ RU DOWHUQDWLYHO\ 0 FDEOH JODQG 7KUHDGHG FRYHU (QFORVXUH 5HWDLQLQJ FROODU 3URFHVV FRQQHFWLRQ 6HQVRU UHIHUHQFH SRLQW Dimensions in mm (inch) 196 SITRANS LR250 with mA/HART Operating Instructions, 10/2019, A5E3222
Dimension drawings 12.14 Hygienic encapsulated antenna (DN50 nozzle/slotted nut to DIN 11851) 12.14 Hygienic encapsulated antenna (DN50 nozzle/slotted nut to DIN 11851) 137 FDEOH HQWU\ RU DOWHUQDWLYHO\ 0 FDEOH JODQG 7KUHDGHG FRYHU (QFORVXUH 5HWDLQLQJ FROODU 7KUHDGHG QXW 6HQVRU UHIHUHQFH SRLQW Dimensions in mm (inch) Note Cut out of process connection and placemen
Dimension drawings 12.15 Hygienic encapsulated antenna (DN80 nozzle/slotted nut to DIN 11851) 12.15 Hygienic encapsulated antenna (DN80 nozzle/slotted nut to DIN 11851) 137 FDEOH HQWU\ RU DOWHUQDWLYHO\ 0 FDEOH JODQG 7KUHDGHG FRYHU (QFORVXUH 5HWDLQLQJ FROODU 7KUHDGHG QXW 6HQVRU UHIHUHQFH SRLQW Dimensions in mm (inch) Note Cut out of process connection and placement of
Dimension drawings 12.16 Hygienic encapsulated antenna (DN100 nozzle/slotted nut to DIN 11851) 12.16 Hygienic encapsulated antenna (DN100 nozzle/slotted nut to DIN 11851) 137 FDEOH HQWU\ RU DOWHUQDWLYHO\ 0 FDEOH JODQG 7KUHDGHG FRYHU (QFORVXUH 5HWDLQLQJ FROODU 7KUHDGHG QXW 6HQVRU UHIHUHQFH SRLQW Dimensions in mm (inch) Note Cut out of process connection and placeme
Dimension drawings 12.17 Hygienic encapsulated antenna (DN50 aseptic slotted nut to DIN 11864-1) 12.17 Hygienic encapsulated antenna (DN50 aseptic slotted nut to DIN 11864-1) 137 FDEOH HQWU\ RU DOWHUQDWLYHO\ 0 FDEOH JODQG 7KUHDGHG FRYHU (QFORVXUH 5HWDLQLQJ FROODU 7KUHDGHG QXW 6HQVRU UHIHUHQFH SRLQW Dimensions in mm (inch) Note Cut out of process connection and placem
Dimension drawings 12.18 Hygienic encapsulated antenna (DN80 aseptic slotted nut to DIN 11864-1) 12.18 Hygienic encapsulated antenna (DN80 aseptic slotted nut to DIN 11864-1) 137 FDEOH HQWU\ RU DOWHUQDWLYHO\ 0 FDEOH JODQG 7KUHDGHG FRYHU (QFORVXUH 5HWDLQLQJ FROODU 7KUHDGHG QXW 6HQVRU UHIHUHQFH SRLQW Dimensions in mm (inch) Note Cut out of process connection and plac
Dimension drawings 12.19 Hygienic encapsulated antenna (DN100 aseptic slotted nut to DIN 11864-1) 12.19 Hygienic encapsulated antenna (DN100 aseptic slotted nut to DIN 11864-1) 137 FDEOH HQWU\ RU DOWHUQDWLYHO\ 0 FDEOH JODQG 7KUHDGHG FRYHU (QFORVXUH 5HWDLQLQJ FROODU 7KUHDGHG QXW 6HQVRU UHIHUHQFH SRLQW Dimensions in mm (inch) Note Cut out of process connection and pl
Dimension drawings 12.20 Hygienic encapsulated antenna (DN50 aseptic flange to DIN 11864-2) 12.20 Hygienic encapsulated antenna (DN50 aseptic flange to DIN 11864-2) 137 FDEOH HQWU\ RU DOWHUQDWLYHO\ 0 FDEOH JODQG 7KUHDGHG FRYHU (QFORVXUH 5HWDLQLQJ FROODU )ODQJH 6HQVRU UHIHUHQFH SRLQW Dimensions in mm (inch) Note Cut out of process connection and flange are show
Dimension drawings 12.21 Hygienic encapsulated antenna (DN80 aseptic flange to DIN 11864-2) 12.21 Hygienic encapsulated antenna (DN80 aseptic flange to DIN 11864-2) 137 FDEOH HQWU\ RU DOWHUQDWLYHO\ 0 FDEOH JODQG 7KUHDGHG FRYHU (QFORVXUH 5HWDLQLQJ FROODU )ODQJH 6HQVRU UHIHUHQFH SRLQW Dimensions in mm (inch) Note Cut out of process connection and flange are sho
Dimension drawings 12.22 Hygienic encapsulated antenna (DN100 aseptic flange to DIN 11864-2) 12.22 Hygienic encapsulated antenna (DN100 aseptic flange to DIN 11864-2) 137 FDEOH HQWU\ RU DOWHUQDWLYHO\ 0 FDEOH JODQG 7KUHDGHG FRYHU (QFORVXUH 5HWDLQLQJ FROODU )ODQJH 6HQVRU UHIHUHQFH SRLQW Dimensions in mm (inch) Note The cut out of the process connection and th
Dimension drawings 12.23 Hygienic encapsulated antenna (DN50 aseptic clamp to DIN 11864-3) 12.23 Hygienic encapsulated antenna (DN50 aseptic clamp to DIN 11864-3) 137 FDEOH HQWU\ RU DOWHUQDWLYHO\ 0 FDEOH JODQG 7KUHDGHG FRYHU (QFORVXUH 5HWDLQLQJ FROODU 3URFHVV FRQQHFWLRQ 6HQVRU UHIHUHQFH SRLQW Dimensions in mm (inch) Note Cut out of process connection is shown
Dimension drawings 12.24 Hygienic encapsulated antenna (DN80 aseptic clamp to DIN 11864-3) 12.24 Hygienic encapsulated antenna (DN80 aseptic clamp to DIN 11864-3) 137 FDEOH HQWU\ RU DOWHUQDWLYHO\ 0 FDEOH JODQG 7KUHDGHG FRYHU (QFORVXUH 5HWDLQLQJ FROODU 3URFHVV FRQQHFWLRQ 6HQVRU UHIHUHQFH SRLQW Dimensions in mm (inch) Note Cut out of process connection is sho
Dimension drawings 12.25 Hygienic encapsulated antenna (DN100 aseptic clamp to DIN 11864-3) 12.25 Hygienic encapsulated antenna (DN100 aseptic clamp to DIN 11864-3) 137 FDEOH HQWU\ RU DOWHUQDWLYHO\ 0 FDEOH JODQG 7KUHDGHG FRYHU (QFORVXUH 5HWDLQLQJ FROODU 3URFHVV FRQQHFWLRQ 6HQVRU UHIHUHQFH SRLQW Dimensions in mm (inch) Note Cut out of process connection is s
Dimension drawings 12.26 Hygienic encapsulated antenna (Tuchenhagen Type N) 12.26 Hygienic encapsulated antenna (Tuchenhagen Type N) 137 FDEOH HQWU\ RU DOWHUQDWLYHO\ 0 FDEOH JODQG 7KUHDGHG FRYHU (QFORVXUH 5HWDLQLQJ FROODU 3URFHVV FRQQHFWLRQ 6HQVRU UHIHUHQFH SRLQW Dimensions in mm (inch) SITRANS LR250 with mA/HART Operating Instructions, 10/2019, A5E32220602-AG 209
Dimension drawings 12.27 Hygienic encapsulated antenna (Tuchenhagen Type F) 12.27 Hygienic encapsulated antenna (Tuchenhagen Type F) 137 FDEOH HQWU\ RU DOWHUQDWLYHO\ 0 FDEOH JODQG 7KUHDGHG FRYHU (QFORVXUH 5HWDLQLQJ FROODU 3URFHVV FRQQHFWLRQ 6HQVRU UHIHUHQFH SRLQW Dimensions in mm (inch) 210 SITRANS LR250 with mA/HART Operating Instructions, 10/2019, A5E32220602-AG
Dimension drawings 12.28 Threaded PVDF antenna 12.28 Threaded PVDF antenna PP 137 FDEOH HQWU\ RU DOWHUQDWLYHO\ 0 FDEOH JODQG WKUHDGHG FRYHU PP PP PP HQFORVXUH HOHFWURQLFV PP PP SURFHVV FRQQHFWLRQ WDJ VHQVRU UHIHUHQFH SRLQW PP WKUHDG WKUHDGHG 39') DQWHQQD PP DQWHQQD 2 ' 7KH FRORU RI WKH DQWHQQD PD\ YDU\ Threaded PVDF antenna dimensions Nominal antenna size Antenna O.D.
Dimension drawings 12.29 Threaded connection markings -3dB in the direction of the polarization axis. See Polarization reference point (Page 29) for an illustration. b) c) 12.29 20m when installed in stillpipe. Threaded connection markings With the exception of the threaded PVDF antenna, threaded connection markings are found on the flat face/faces of the process connection.
Dimension drawings 12.30 Raised-Face flange 12.30 Raised-Face flange Stainless steel or optional alloy N06022/2.4602 (Hastelloy® C-22) 2 1 3 4 $ $ 5 6 7 10 9 8 9,(: $ $ ① ② ③ ④ ⑤ ⑥ ⑦ ⑧ ⑨ ⑩ angle of adjacent bolt holes bolt hole diameter bolt hole circle diameter waveguide mounting hole Flange O.D. bolt hole circle diameter facing height facing diameter sensor reference point thickness Raised-Face flange dimensions Flange bolt hole pattern ⑤ ③ ② Flange O.D.
Dimension drawings 12.30 Raised-Face flange Flange bolt hole pattern ⑤ ③ ② Flange O.D. (mm) Bolt hole circle Ø (mm) Bolt hole Ø (mm) DN150 PN10/PN16 285 240 22 DN50 PN25/PN40 165 160 DN80 PN25/PN40 200 DN100 PN25/PN40 235 DN150 PN25/PN40 300 Pipe size No.
Dimension drawings 12.31 Raised-Face flange per EN 1092-1 for flanged encapsulated antenna 12.31 Raised-Face flange per EN 1092-1 for flanged encapsulated antenna Stainless steel 2 1 3 4 “A” “A” 5 6 8 7 ① ② ③ ④ VIEW “A”-"A" angle of adjacent bolt holes bolt hole diameter bolt hole circle diameter antenna SITRANS LR250 with mA/HART Operating Instructions, 10/2019, A5E32220602-AG ⑤ ⑥ ⑦ ⑧ flange O.D.
Dimension drawings 12.31 Raised-Face flange per EN 1092-1 for flanged encapsulated antenna Raised-Face flange dimensions Pipe size Flange class ⑤ ③ ② Flange O.D. [mm (inch)] Bolt hole circle Ø [mm (inch)] Bolt hole Ø [mm (inch)] 152 (5.98) 120.7 (4.75) 19 (0.75) 3" 190 (7.48) 152.4 (6.00) 4" 230 (9.06) 190.5 (7.50) 6" 280 (11.02) 241.3 (9.50) 22.2 (0.87) 155 (6.10) 125 (4.92) 18 (0.71) DN80 200 (7.87) 160 (6.30) DN100 220 (8.66) 180 (7.09) DN150 285 (11.22) 240 (9.45) 22 (0.
Dimension drawings 12.32 Flat-Face flange 12.
Dimension drawings 12.32 Flat-Face flange Flat-Face flange dimensions Flange size a) Flange class Flange O.D. Bolt hole circle Ø Bolt hole Ø No. of bolt holes Thickness 2" ASME 150 lb 6.0" 4.75" 0.75" 4 0.88" 3" ASME 150 lb 7.5" 6.0" 0.75" 4 0.96" 4" ASME 150 lb 9.0" 7.50" 0.75" 8 1.25" 2" ASME 300 lb 6.50" 5.00" 0.75" 8 1.12" 3" ASME 300 lb 8.25" 6.62" 0.88" 8 1.38" 4" ASME 300 lb 10.00" 7.88" 0.88" 8 1.50" DN50 EN PN16 165 mm 125 mm 18 mm 4 24.
Dimension drawings 12.33 Aseptic/hygienic flange DN50, DN80, DN100 for DIN 11864-2 12.
Dimension drawings 12.
Dimension drawings 12.34 Process connection tag (pressure rated versions) Aseptic/hygienic flange DN100 $ $ 9,(: $ $ Dimensions in mm (inch) 12.
Dimension drawings 12.34 Process connection tag (pressure rated versions) Item Sample Text Comments/Explanation DESIGN TEMP. (TS) 200 ºC Maximum Allowable Working Temperature MINIMUM PROCESS 15.9 BAR AT 40 ºC Minimum Wetted Process Conditions TEST PRESSURE (PT) 22.7 BAR Production Test Pressure TEST DATE 10/11/11 Date of Pressure Test (Year/Month/Day) CONNECTION SERIES ASME B16.5 Flange Series: dimensional pattern based on ASME B16.
Appendix A: Technical reference A Note Where a number follows the parameter name [for example, Master Reset (4.1.)] this is the parameter access number via the handheld programmer. See Parameter Reference (Page 107) for a complete list of parameters. A.1 Principles of operation SITRANS LR250 is a 2-wire 25 GHz pulse radar level transmitter for continuous monitoring of liquids and slurries. (The microwave output level is significantly less than that emitted from cellular phones.
Appendix A: Technical reference A.2 Echo Processing Process Intelligence is able to differentiate between the true microwave reflections from the surface of the material and unwanted reflections being returned from obstructions such as seam welds or supports within a vessel. The result is repeatable, fast and reliable measurement. This technology was developed as result of field data gained over some twenty years from more than 1,000,000 installations in many industries around the world.
Appendix A: Technical reference A.2 Echo Processing The device characterizes all echoes that rise above the TVT as potential good echoes. Each peak is assigned a rating based on its strength, area, height above the TVT, and reliability, amongst other characteristics. Algorithm (2.8.4.1.) The true echo is selected based on the setting for the Echo selection algorithm. Options are true First Echo, Largest Echo, or best of First and Largest. Position Detect (2.8.4.2.
Appendix A: Technical reference A.2 Echo Processing 1 2 3 4 ① ② ③ ④ default TVT material echo vessel bottom echo selected echo marker Example: CLEF enabled Vessel height: 1.5 m; CLEF range set to 0.
Appendix A: Technical reference A.2 Echo Processing A.2.3 CLEF Range CLEF Range (2.8.4.4.) is referenced from Low Calibration Point (process empty level ). When the Hybrid algorithm is selected in Position Detect (2.8.4.2.), the CLEF algorithm will be applied up to the limit of CLEF Range. Above this limit the Center algorithm will be applied. 4 3 2 1 ① ② ③ ④ ⑤ A.2.
Appendix A: Technical reference A.2 Echo Processing A.2.6 Auto False Echo Suppression Note ● For detailed instructions on using this feature via PDM see Auto False Echo Suppression (Page 74). ● For detailed instructions on using this feature via the handheld programmer see Auto False Echo Suppression (2.8.7.1.).
Appendix A: Technical reference A.2 Echo Processing Example after Auto False Echo Suppression 1 /HYHO GE 5 4 2 ① ② ③ A.2.7 Auto False Echo Suppression Range set to 2 m 3 6 learned TVT false echo Auto False Echo Suppression Range ④ ⑤ ⑥ default TVT material echo echo marker Measurement Range Near Range (2.8.1.) Near Range programs SITRANS LR250 to ignore the zone in front of the antenna. The default blanking distance is 50 mm (1.97") from the end of the antenna.
Appendix A: Technical reference A.3 Analog Output Once the real process fill/empty rate (m/s by default) is established, a response rate can be selected that is slightly higher than the application rate. Response Rate automatically adjusts the filters that affect the output response rate. Response Rate (2.4.1.) * A.2.9 Fill Rate per Minute (2.4.2.)/Empty Rate per Mi‐ nute (2.4.3.) Slow Damping Filter (2.2.4.) 0.1 m/min (0.32 ft/min) 10 s Medium 1.0 m/min (3.28 ft.min) 10 s Fast 10.0 m/min (32.
Appendix A: Technical reference A.3 Analog Output A.3.1 Sensor Mode This parameter controls the input. Depending on the reference point used, the measurement reports either Level, Space, or Distance. By default Sensor Mode is set to Level. Operation Description NO SERVICE 1) Reference point Measurement and associated loop current not being updated. Device defaults to Fail-safe mode 1).
Appendix A: Technical reference A.3 Analog Output A.3.3 Loss of Echo (LOE) A loss of echo (LOE) occurs when the calculated measurement is judged to be unreliable because the echo confidence value has dropped below the echo confidence threshold. Confidence (2.8.6.1.) describes the quality of an echo. Higher values represent higher quality. Echo Threshold (2.8.4.3.) defines the minimum confidence value required for an echo to be accepted as valid and evaluated.
Appendix A: Technical reference A.4 Maximum Process Temperature Chart A.4 Maximum Process Temperature Chart WARNING Exceeded maximum internal and process temperatures Danger of device malfunction ● Internal temperature must not exceed +80 ° C (+176 °F). ● Process temperature must not exceed limits specified by the antenna type. Note ● The chart below is for guidance only. ● The chart does not represent every possible process connection arrangement.
Appendix A: Technical reference A.5 Process Pressure/Temperature derating curves A.5 Process Pressure/Temperature derating curves WARNING Exceeded maximum permissible operating pressure Danger of injury or poisoning. The maximum permissible operating pressure depends on the device version. The device can be damaged if the operating pressure is exceeded. Hot, toxic and corrosive process media could be released.
Appendix A: Technical reference A.5 Process Pressure/Temperature derating curves A.5.1 Pressure Equipment Directive, PED, 14/68/EU Siemens Level Transmitters with flanged, threaded, or sanitary clamp type process mounts have no pressure-bearing housing of their own and, therefore, do not come under the Pressure Equipment Directive as pressure or safety accessories (see EU Commission Guideline A-08 and A-20). A.5.2 Stainless steel horn antenna 1.
Appendix A: Technical reference A.5 Process Pressure/Temperature derating curves A.5.3 Stainless steel flanged horn antenna JIS B 2220, 10K: 50A, 80A, and 100A P 30 bar (435 psi) 20 bar (290 psi) 10 bar (145 psi) 1 -40°F (-40°C) 14°F (-10°C) 100°F (38°C) 200°F (93°C) 300°F (149°C) 400°F (204°C) T -1 bar (-14.
Appendix A: Technical reference A.5 Process Pressure/Temperature derating curves ① P T Atmospheric Allowable operating pressures Allowable operating temperatures EN1092-1, PN40: DN50, DN80, DN100, and DN150 P 40 bar (580 psi) 30 bar (435 psi) PN40 20 bar (290 psi) 10 bar (145 psi) 1 -40°F (-40°C) 14°F (-10°C) 100°F (38°C) 200°F (93°C) 300°F (149°C) 400°F (204°C) T -1 bar (-14.
Appendix A: Technical reference A.5 Process Pressure/Temperature derating curves ASME B16.5, Class 150: 2", 3", and 4" NPS P 40 bar (580 psi) 30 bar (435 psi) 20 bar (290 psi) ASME 150# 10 bar (145 psi) 1 -40°F (-40°C) 14°F (-10°C) 100°F (38°C) 200°F (93°C) 300°F (149°C) 400°F (204°C) T -1 bar (-14.
Appendix A: Technical reference A.5 Process Pressure/Temperature derating curves A.5.4 PTFE flanged encapsulated antenna ASME B16.
Appendix A: Technical reference A.5 Process Pressure/Temperature derating curves EN1092-1, PN10/16: DN50, DN80, DN100, and DN150 3 DN150 30 bar (435 psi) DN100 DN50, DN80 20 bar (290 psi) 31 10 bar (145 psi) 1 r) r) r& r& r) r& r) r& r) r& r) r& -1 bar (-14.
Appendix A: Technical reference A.5 Process Pressure/Temperature derating curves A.5.
Appendix A: Technical reference A.5 Process Pressure/Temperature derating curves A.5.6 PVDF antenna ASME B1.20.1 2" NPT, EN ISO 228-1 2" G (BSPP), EN 10226-1 2" R (BSPT) P 5 bar (73.5 psi) 4 bar (58.0 psi) 3 bar (4.35 psi) 2 bar (2.90 psi) 1 bar (14.5 psi) 1 -40°F (-40°C) 14°F (-10°C) 100°F (38°C) 200°F (93°C) T -1 bar (-14.5 psi) RT -50°C (-58°F) 0°C (32°F) 50°C (122°F) -29°C (-20°F) ① P T A.5.
Appendix A: Technical reference A.
Appendix A: Technical reference A.6 Loop power Note For pressure applications, all attachment hardware must be suitably rated.
Appendix A: Technical reference A.6 Loop power 3 1 2 5 4 ① ② ③ ④ ⑤ A.6.1 Loop voltage VL Power supply Loop resistance RL Loop current IL LR250 Allowable operating area of SITRANS LR250 Note The curves below apply to a standalone device, configured via the Siemens handheld programmer. A.6.2 Curve 1 (General Purpose, Intrinsically Safe, Non-Sparking, Non-incendive) Loop Voltage versus Loop Resistance /RRS 5HVLVWDQFH 5 / 3RZHU 6XSSO\ 5HTXLUHPHQWV 9 ˖ 9 ˖ 9 ˖ $//2:$%/( 23(5$7,1*
Appendix A: Technical reference A.7 Startup behavior Note When using HART communications, the minimum voltage with 220 Ohms (RL) is 16.3 V DC. A.6.3 Curve 2 (Flameproof, Increased Safety, Explosion-proof) Loop Voltage versus Loop Resistance 3RZHU 6XSSO\ 5HTXLUHPHQWV /RRS 5HVLVWDQFH 5 / 9 ˖ 9 ˖ 9 ˖ 5/ 9/ 9 ˖ $//2:$%/( 23(5$7,1* $5($ 5/ 9/ /RRS 9ROWDJH 9/ Note When using HART communications, the minimum voltage with 220 Ohms (RL) is 20.
Appendix B: HART communications B Highway Addressable Remote Transducer, HART, is an industrial protocol that is superimposed on the 4-20 mA signal. It is an open standard, and full details about HART can be obtained from the HART Communication Foundation website: HART Communication Foundation (http://www.hartcomm.org/) The radar device can be configured over the HART network using either the HART Communicator 375 by Fisher-Rosemount, or a software package.
Appendix B: HART communications B.
Appendix B: HART communications B.
Appendix B: HART communications B.5 HART version 6(59,&( ,17(59$/ 727$/ 6(59,&( ,17(59$/ 7,0( (/$36(' 6,1&( /$67 6(59,&( 5(0$,1,1* /,)(7,0( 0$,17(1$1&( 5(48,5(' /,0,7 0$,17(1$1&( '(0$1'(' /,0,7 0$,17(1$1&( $/(57 $&7,9$7,21 6(59,&( $&.
Appendix C: Certificates and support C.1 C Technical support Technical support If this documentation does not provide complete answers to any technical questions you may have, contact Technical Support at: ● Support request (http://www.siemens.com/automation/support-request) ● More information about our Technical Support is available at Technical Support (http://www.siemens.
Appendix C: Certificates and support C.
13 List of abbreviations Short form Long form 3-A 3-A Sanitary Standards, Inc.
List of abbreviations 254 SITRANS LR250 with mA/HART Operating Instructions, 10/2019, A5E32220602-AG
LCD menu structure 14.1 14 LCD menu structure Note ● In Navigation mode, ARROW keys ( ) navigate the menu in the direction of the arrow. See Parameter Reference (Page 107) for detailed information and instructions.
LCD menu structure 14.1 LCD menu structure /(9(/ 0(7(5 48,&. 67$57 :,= 48,&.
LCD menu structure 14.1 LCD menu structure 6(783 FRQW G 6,*1$/ 352& 1($5 5$1*( )$5 5$1*( 3523$* )$&725 (&+2 6(/(&7 $/*25,7+0 326 '(7(&7 (&+2 7+5(6+2/' &/() 5$1*( 6$03/,1* (&+2 /2&.
LCD menu structure 14.1 LCD menu structure 6(59,&( FRQW G 6(59,&( 6&+(' 6(59 ,17(59$/ 7,0( /$67 6(59 7,0( 1(;7 6(59 5(0,1'(5 $&7,9 5(0,1' 5(4 5(0,1' '(0 0$,17 67$7 $&. 67$786 $&.
Glossary accuracy degree of conformity of a measure to a standard or a true value. algorithm a prescribed set of well-defined rules or processes for the solution of a problem in a finite number of steps. ambient temperature the temperature of the surrounding air that comes in contact with the enclosure of the device. antenna an aerial which sends out and receives a signal in a specific direction. There are four basic types of antenna in radar level measurement, horn, parabolic, rod, and waveguide.
Glossary blanking a blind zone extending away from the reference point plus any additional shield length.The device is programmed to ignore this zone. capacitance the property of a system of conductors and dielectrics that permits the storage of electricity when potential differences exist between the conductors. Its value is expressed as the ratio of a quantity of electricity to a potential difference, and the unit is a Farad. confidence see Echo Confidence.
Glossary Echo Confidence describes the quality of an echo. Higher values represent higher quality. Echo Threshold defines the minimum value required for an echo to be accepted as valid and evaluated. Echo Lock Window a window centered on an echo in order to locate and display the echo’s position and true reading. Echoes outside the window are not immediately processed. Echo Marker a marker that points to the processed echo. Echo Processing the process by which the radar unit determines echoes.
Glossary inductance the property of an electric circuit by virtue of which a varying current induces an electromotive force in that circuit or in a neighboring circuit. The unit is a Henry. microwaves the term for the electromagnetic frequencies occupying the portion of the radio frequency spectrum from 1 GHz to 300 GHz. multiple echoes secondary echoes that appear as double, triple, or quadruple echoes in the distance from the target echo. Near Blanking see Blanking.
Glossary radar radar is an acronym for RAdio Detection And Ranging. A device that radiates electromagnetic waves and utilizes the reflection of such waves from distant objects to determine their existence or position. range distance between a transmitter and a target. range extension the distance below the zero percent or empty point in a vessel.
Glossary stillpipe a pipe that is mounted inside a vessel parallel to the vessel wall, and is open to the vessel at the bottom. TVT (Time Varying Threshold) a time-varying curve that determines the threshold level above which echoes are determined to be valid. two wire radar a low-energy radar. Can be loop powered, analog, intrinsically safe 4 to 20 mA, or a digital (BUS) transmitter. waveguide antenna a hollow, metallic tube that transmits a microwave signal to the product target.
Index A C Abbreviations and identifications list, 253 access control remote access, 147 activating, 46 agitator blade detection avoiding, 124 AMS Device Manager features, 84 analog output explanation, 230 fail signal, 167 select, 78 signal range, 167 simulate via PDM, 78 analog output scale setup, 114 antenna replacement, 155 antenna types flanged encapsulated antenna, 190, 192 flanged horn antenna, 186 flanged horn antenna with extension, 189 threaded horn, 181 threaded PVDF antenna, 211 Auto False Echo
Index threaded horn with extension, 182, 185 threaded PVDF antenna, 211 Disassembly, 35 Disposal, 157 Documentation, 251 E echo confidence parameter setup, 125 echo processing Process Intelligence, 223 Echo Profile data logging, 73 view via LUI, 56 view via PDM, 72 Echo Profile Utilities Auto False Echo Suppression, 71 Echo Profile, 71 TVT Shaper, 71 echo selection Algorithm, 225 CLEF (Constrained Leading Edge Fit), 225 Position algorithm, 225 time varying threshold (TVT), 224 Echo Setup quick access via
Index L Language, 148 LCD display contrast adjustment, 146 echo profile viewing, 56 fast mode, 145 measurement mode, 47 lens replacement, 155 lid-lock set screw, 38 linearization via AMS, 88 Lithium batteries Safety, 45 Local User Interface (LUI), 47 LOE Fail-safe Mode, 232 loop test simulate analog output, 78 loop voltage vs.
Index programmer handheld, 49 programming adjust parameters via PDM, 70 via the handheld programmer, 50 propogation factor values, 121 Q Qualified personnel, 17 Quick Start Wizard via LUI, 53 via SIMATIC PDM, 66 R raised face flange markings, 213, 215 reading erratic troubleshooting, 164 reading incorrect troubleshooting, 164 repair cautions, 155 excluded liability, 155 reset see Master Reset, 131 Response Rate explanation, 229 Return procedure, 157 S Scan device synchronize parameters via AMS, 86 Scope
Index troubleshooting communication, 159 operation, 163 TVT (time varying threshold) explanation, 224 TVT Shaper manual adjustment via PDM, 73 via PDM, 71 U unlock value local access, 147 V vessel shape selection, 116 W Warranty, 12 Wear powered hours, 79 poweron resets, 79 view via PDM, 79 wiring cables, 38 connecting HART, 39 hazardous areas, 39 write protect local access, 148 SITRANS LR250 with mA/HART Operating Instructions, 10/2019, A5E32220602-AG 269
Index 270 SITRANS LR250 with mA/HART Operating Instructions, 10/2019, A5E32220602-AG