Operating Instructions
Table Of Contents
- SITRANS LR250 with mA/HART
- Legal information - Warning notice system
- Table of contents
- 1 Introduction
- 2 Safety information
- 3 Description
- 4 Installing/mounting
- 5 Connecting
- 6 Commissioning
- 7 Remote operation
- 7.1 Operating via SIMATIC PDM
- 7.1.1 Functions in SIMATIC PDM
- 7.1.2 Initial setup
- 7.1.3 Updating the Electronic Device Description (EDD)
- 7.1.4 Quick Start Wizard via SIMATIC PDM
- 7.1.5 Changing parameter settings using SIMATIC PDM
- 7.1.6 Parameters accessed via pull-down menus
- 7.1.6.1 Echo profile utilities
- 7.1.6.2 Echo profile
- 7.1.6.3 View saved echo profiles
- 7.1.6.4 TVT Shaper
- 7.1.6.5 Auto false echo suppression
- 7.1.6.6 Echo setup
- 7.1.6.7 Maintenance
- 7.1.6.8 Select analog output
- 7.1.6.9 Self test
- 7.1.6.10 Loop test
- 7.1.6.11 Configuration flag reset
- 7.1.6.12 Master reset
- 7.1.6.13 Wear
- 7.1.6.14 HART Communication
- 7.1.6.15 Process variables
- 7.1.6.16 Trend
- 7.1.6.17 Device status
- 7.1.6.18 Update
- 7.1.6.19 Security
- 7.2 Operating via AMS Device Manager
- 7.3 Operating via FDT (Field Device Tool)
- 7.1 Operating via SIMATIC PDM
- 8 Parameter reference
- 9 Service and maintenance
- 10 Diagnosing and troubleshooting
- 11 Technical data
- 12 Dimension drawings
- 12.1 Universal polymeric flange
- 12.2 Aluminum horn antenna
- 12.3 Mounting bracket
- 12.4 Aluminum horn with polypropylene lens antenna and polymeric flange
- 12.5 Stainless steel threaded horn antenna
- 12.6 Stainless steel threaded horn antenna with extension
- 12.7 Stainless steel flanged horn antenna
- 12.8 Stainless steel flanged horn antenna with extension
- 12.9 PTFE Flanged encapsulated antenna (2"/DN50/50A sizes only)
- 12.10 PTFE Flanged encapsulated antenna (3"/DN80/80A sizes and larger)
- 12.11 Hygienic encapsulated antenna (2" ISO 2852 sanitary clamp)
- 12.12 Hygienic encapsulated antenna (3" ISO 2852 sanitary clamp)
- 12.13 Hygienic encapsulated antenna (4" ISO 2852 sanitary clamp)
- 12.14 Hygienic encapsulated antenna (DN50 nozzle/slotted nut to DIN 11851)
- 12.15 Hygienic encapsulated antenna (DN80 nozzle/slotted nut to DIN 11851)
- 12.16 Hygienic encapsulated antenna (DN100 nozzle/slotted nut to DIN 11851)
- 12.17 Hygienic encapsulated antenna (DN50 aseptic slotted nut to DIN 11864-1)
- 12.18 Hygienic encapsulated antenna (DN80 aseptic slotted nut to DIN 11864-1)
- 12.19 Hygienic encapsulated antenna (DN100 aseptic slotted nut to DIN 11864-1)
- 12.20 Hygienic encapsulated antenna (DN50 aseptic flange to DIN 11864-2)
- 12.21 Hygienic encapsulated antenna (DN80 aseptic flange to DIN 11864-2)
- 12.22 Hygienic encapsulated antenna (DN100 aseptic flange to DIN 11864-2)
- 12.23 Hygienic encapsulated antenna (DN50 aseptic clamp to DIN 11864-3)
- 12.24 Hygienic encapsulated antenna (DN80 aseptic clamp to DIN 11864-3)
- 12.25 Hygienic encapsulated antenna (DN100 aseptic clamp to DIN 11864-3)
- 12.26 Hygienic encapsulated antenna (Tuchenhagen Type N)
- 12.27 Hygienic encapsulated antenna (Tuchenhagen Type F)
- 12.28 Threaded PVDF antenna
- 12.29 Threaded connection markings
- 12.30 Raised-Face flange
- 12.31 Raised-Face flange per EN 1092-1 for flanged encapsulated antenna
- 12.32 Flat-Face flange
- 12.33 Aseptic/hygienic flange DN50, DN80, DN100 for DIN 11864-2
- 12.34 Process connection tag (pressure rated versions)
- A Appendix A: Technical reference
- A.1 Principles of operation
- A.2 Echo Processing
- A.3 Analog Output
- A.4 Maximum Process Temperature Chart
- A.5 Process Pressure/Temperature derating curves
- A.5.1 Pressure Equipment Directive, PED, 14/68/EU
- A.5.2 Stainless steel horn antenna
- A.5.3 Stainless steel flanged horn antenna
- A.5.4 PTFE flanged encapsulated antenna
- A.5.5 Aluminum horn, with polypropylene lens antenna version with polypropylene flange
- A.5.6 PVDF antenna
- A.5.7 Hygienic encapsulated antenna
- A.6 Loop power
- A.7 Startup behavior
- B Appendix B: HART communications
- C Appendix C: Certificates and support
- 13 List of abbreviations
- 14 LCD menu structure
- Glossary
- Index
User-configurable recommended time between product inspections.
Values Units
a)
: hours, days, years
Range: 0 to 20 years
Default: 1.0 year
a)
Units are selectable only via SIMATIC PDM.
Time Since Last Service (4.4.2.)
Time elapsed since last service. Can be reset to zero after performing a service.
To reset to zero:
● In SIMATIC PDM, open the menu Device – Maintenance, click on the Service Schedule tab,
and click on Service Performed to restart the timer and clear any fault messages.
● Via the handheld programmer, manually reset Time since Last Service (4.4.2.) to zero.
Time Until Next Service (4.4.3.)
Read only. Service Interval (4.4.1.) less Time Since Last Service (4.4.2.).
Activation of Reminders (4.4.4.)
Note
To modify this parameter via SIMATIC PDM it must be accessed via the pull-down
menu Device – Maintenance.
Allows you to enable a maintenance reminder.
Options * Timer OFF
ON - no reminders checked
ON - Reminder 1 (Maintenance Required) checked
ON - Reminders 1 and 2 checked
ON - Reminder 2 (Maintenance Demanded) checked
1. First set the values in Reminder 1 before Service (Required) (4.4.5.)/Reminder 2 before
Service (Demanded) (4.4.6.).
2. Select the desired Reminder Activation option.
Reminder 1 before Service (Required) (4.4.5.)
If Time Until Next Service (4.4.3.) is equal to or less than this value, the device generates
a Maintenance Required reminder.
Values Range: 0 to Service Interval (4.4.1.)
Default: 0.164 years
Parameter reference
SITRANS LR250 with mA/HART
140 Operating Instructions, 10/2019, A5E32220602-AG