Operating Instructions
Table Of Contents
- SITRANS LR250 with mA/HART
- Legal information - Warning notice system
- Table of contents
- 1 Introduction
- 2 Safety information
- 3 Description
- 4 Installing/mounting
- 5 Connecting
- 6 Commissioning
- 7 Remote operation
- 7.1 Operating via SIMATIC PDM
- 7.1.1 Functions in SIMATIC PDM
- 7.1.2 Initial setup
- 7.1.3 Updating the Electronic Device Description (EDD)
- 7.1.4 Quick Start Wizard via SIMATIC PDM
- 7.1.5 Changing parameter settings using SIMATIC PDM
- 7.1.6 Parameters accessed via pull-down menus
- 7.1.6.1 Echo profile utilities
- 7.1.6.2 Echo profile
- 7.1.6.3 View saved echo profiles
- 7.1.6.4 TVT Shaper
- 7.1.6.5 Auto false echo suppression
- 7.1.6.6 Echo setup
- 7.1.6.7 Maintenance
- 7.1.6.8 Select analog output
- 7.1.6.9 Self test
- 7.1.6.10 Loop test
- 7.1.6.11 Configuration flag reset
- 7.1.6.12 Master reset
- 7.1.6.13 Wear
- 7.1.6.14 HART Communication
- 7.1.6.15 Process variables
- 7.1.6.16 Trend
- 7.1.6.17 Device status
- 7.1.6.18 Update
- 7.1.6.19 Security
- 7.2 Operating via AMS Device Manager
- 7.3 Operating via FDT (Field Device Tool)
- 7.1 Operating via SIMATIC PDM
- 8 Parameter reference
- 9 Service and maintenance
- 10 Diagnosing and troubleshooting
- 11 Technical data
- 12 Dimension drawings
- 12.1 Universal polymeric flange
- 12.2 Aluminum horn antenna
- 12.3 Mounting bracket
- 12.4 Aluminum horn with polypropylene lens antenna and polymeric flange
- 12.5 Stainless steel threaded horn antenna
- 12.6 Stainless steel threaded horn antenna with extension
- 12.7 Stainless steel flanged horn antenna
- 12.8 Stainless steel flanged horn antenna with extension
- 12.9 PTFE Flanged encapsulated antenna (2"/DN50/50A sizes only)
- 12.10 PTFE Flanged encapsulated antenna (3"/DN80/80A sizes and larger)
- 12.11 Hygienic encapsulated antenna (2" ISO 2852 sanitary clamp)
- 12.12 Hygienic encapsulated antenna (3" ISO 2852 sanitary clamp)
- 12.13 Hygienic encapsulated antenna (4" ISO 2852 sanitary clamp)
- 12.14 Hygienic encapsulated antenna (DN50 nozzle/slotted nut to DIN 11851)
- 12.15 Hygienic encapsulated antenna (DN80 nozzle/slotted nut to DIN 11851)
- 12.16 Hygienic encapsulated antenna (DN100 nozzle/slotted nut to DIN 11851)
- 12.17 Hygienic encapsulated antenna (DN50 aseptic slotted nut to DIN 11864-1)
- 12.18 Hygienic encapsulated antenna (DN80 aseptic slotted nut to DIN 11864-1)
- 12.19 Hygienic encapsulated antenna (DN100 aseptic slotted nut to DIN 11864-1)
- 12.20 Hygienic encapsulated antenna (DN50 aseptic flange to DIN 11864-2)
- 12.21 Hygienic encapsulated antenna (DN80 aseptic flange to DIN 11864-2)
- 12.22 Hygienic encapsulated antenna (DN100 aseptic flange to DIN 11864-2)
- 12.23 Hygienic encapsulated antenna (DN50 aseptic clamp to DIN 11864-3)
- 12.24 Hygienic encapsulated antenna (DN80 aseptic clamp to DIN 11864-3)
- 12.25 Hygienic encapsulated antenna (DN100 aseptic clamp to DIN 11864-3)
- 12.26 Hygienic encapsulated antenna (Tuchenhagen Type N)
- 12.27 Hygienic encapsulated antenna (Tuchenhagen Type F)
- 12.28 Threaded PVDF antenna
- 12.29 Threaded connection markings
- 12.30 Raised-Face flange
- 12.31 Raised-Face flange per EN 1092-1 for flanged encapsulated antenna
- 12.32 Flat-Face flange
- 12.33 Aseptic/hygienic flange DN50, DN80, DN100 for DIN 11864-2
- 12.34 Process connection tag (pressure rated versions)
- A Appendix A: Technical reference
- A.1 Principles of operation
- A.2 Echo Processing
- A.3 Analog Output
- A.4 Maximum Process Temperature Chart
- A.5 Process Pressure/Temperature derating curves
- A.5.1 Pressure Equipment Directive, PED, 14/68/EU
- A.5.2 Stainless steel horn antenna
- A.5.3 Stainless steel flanged horn antenna
- A.5.4 PTFE flanged encapsulated antenna
- A.5.5 Aluminum horn, with polypropylene lens antenna version with polypropylene flange
- A.5.6 PVDF antenna
- A.5.7 Hygienic encapsulated antenna
- A.6 Loop power
- A.7 Startup behavior
- B Appendix B: HART communications
- C Appendix C: Certificates and support
- 13 List of abbreviations
- 14 LCD menu structure
- Glossary
- Index
Material
Selects the appropriate echo processing algorithms for the material [see Position Detect
(2.8.4.2.) for more detail].
Options * LIQUID
LIQUID LOW DK
a)
(low dielectric liquid – CLEF algorithm enabled)
a)
dK < 3.0
Response Rate
Sets the reaction speed of the device to measurement changes in the target range. Use a
setting just faster than the maximum filling or emptying rate (whichever is greater).
Options Response Rate (1.3.) Fill rate per Minute (2.4.2.)/Empty rate per Mi‐
nute (2.4.3.)
* SLOW 0.1 m/min (0.32 ft/min)
MED 1.0 m/min (3.28 ft/min)
FAST 10.0 m/min (32.8 ft/min)
Units
Sensor measurement units.
Commissioning
6.2 Operating via the handheld programmer
SITRANS LR250 with mA/HART
54 Operating Instructions, 10/2019, A5E32220602-AG