6.3.2 Horizontal vessel with volume measurement
Note
The minimum distance from the flange face to the target is limited by Near Range (2.8.1.).
To obtain level measurement/4 to 20 mA output proportional to vessel volume in a chemical
vessel:
Low Calibration Point = 3.5 m (11.48 ft) from sensor reference point
High Calibration Point = 0.5 m (1.64 ft) from sensor reference point
Max. fill/empty rate = 0.2 m/min (0.65 ft/min)
Select vessel shape, Parabolic Ends, and enter values for A and L, to obtain a volume reading
instead of level.
In the event of a loss of echo: SITRANS LR250 is to go into Fail-safe High after 2 minutes.
1
3
4
5
6 6
8
9
2
7
Sensor reference point A = 0.8 m
0.5 m L = 6 m
High Calibration Point 100% = 8000 L
3.5. m Volume reading
Low Calibration Point
Parameter type Parameter No./Name Options/
Values
Function
Quick Start Wiz‐
ard parameters
Introduction NEXT Continue with Wizard.
Language NEXT Continue with current lan‐
guage.
Material LIQUID
Response Rate MED Medium =1 m/minute
Units M meters
Operation LEVEL Level is reported as Vol‐
ume when a vessel
shape is selected.
Low Calibration Point 3.5 3.5 m (11.48 ft)
High Calibration Point 0.5 0.5 m (1.64 ft)
Wizard Complete FINISH Transfers Quick Start set‐
tings to device.
Commissioning