Parameter type Parameter No./Name Options/Values Function
Quick Start Wizard Introduction NEXT Continue with Wizard.
Language NEXT Continue with current
language.
Material LIQUID LOW DK
Response Rate MED Medium =1 m/minute
Units M meters
Operation LEVEL Level is reported as Vol‐
ume when a vessel
shape is selected.
Low Calibration Point 5 5 m (16.5 ft)
High Calibration Point 0.5 0.5 m (1.64 ft)
Wizard Complete FINISH Transfers Quick Start
settings to device.
Independent parame‐
ters
Propagation Factor
(2.8.3.)
a)
0.988 P.F. for a 50 mm (1.96")
I.D. stillpipe
Position Detect
(2.8.4.2.)
HYBRID
CLEF Range (2.8.4.4.)
a)
4.3 Low calibration point -
0.7 m = 4.3 m (14.1 ft)
a)
The recommended values for the propagation factor and for CLEF range are dependent on the
stillpipe diameter. Refer to the next table for values.
Propagation Factor/Stillpipe Diameter
1
2
3
4
$
sensor reference point low calibration point
air gap CLEF range 2.8.4.4.
A 700 or 1000 mm (see CLEF Range settings in table below)
Values Range 0.3 to 1.0 depending on pipe size
Default 1.0000
Nominal Pipe Size
a)
40 mm (1.5") 50 mm (2") 80 mm (3") 100 mm (4")
Commissioning
6.3 Application examples
SITRANS LR250 with mA/HART
Operating Instructions, 10/2019, A5E32220602-AG 61