Operating Instructions

Table Of Contents
7.1.6.2 Echo profile
Note
Double click on each axis to see the Xscale and Data Scale values.
To zoom in to a section of the profile, left-click and drag a marquee around it. Right click
inside the window to zoom out.
Expand or compress the x and/or y axes:
Left-click on the axis and drag in either direction to reposition the low end of the scale.
Right-click on the axis and drag in either direction to reposition the high end of the scale.
After saving a profile click on OK, not the x button, to close the Echo Profile Utilities window,
otherwise the profile will not be saved.
In the Echo Profile Utilities window click the Echo Profile tab.
Initial profile graph is blank upon entry to dialog. Click Measure to update the profile.
It is recommended to use the Detailed resolution view of the echo profile for troubleshooting.
For faster and more coarse views, the Standard resolution may be used.
Click Save and in the new window enter a name and click OK.
Click OK to exit.
1
2
Resolution
Echo Profile Time Based Storage
7.1.6.3 View saved echo profiles
To view a saved profile, click on the tab View Saved Echo Profiles.
Remote operation
7.1 Operating via SIMATIC PDM
SITRANS LR250 with mA/HART
72 Operating Instructions, 10/2019, A5E32220602-AG