7.2.1.8 Quick Start Wizard via AMS Device Manager
Note
A reset to Factory Defaults should be performed before running the Quick Start Wizard if
device has been used in a previous application. See Master Reset (Page 85).
The layout of the dialog boxes shown may vary according to the resolution setting for your
computer monitor.
At each step, you can accept the default values without modification and click on the next
step to proceed.
After modifying parameters click on Apply inside the Quick Start window to write the new
values to the device.
Click on OK only if you wish to update all parameters to the device and close AMS.
A Wizard provides an easy 5-step Quick Start procedure that configures the device for a simple
application.
Quick Start
Step 1 – Identification
1. Navigate to Configure/Setup > Wizard > Quick Start.
2. Click on Step 1.
3. You can accept the default values without modification. (Description, Message, and
Installation Date fields can be left blank.) If desired, make changes then click on Apply.
Remote operation
7.2 Operating via AMS Device Manager
SITRANS LR250 with mA/HART
Operating Instructions, 10/2019, A5E32220602-AG 87