Breakpoints
Maximum volume
1. Click on the Breakpoints tab and enter values for level and volume for up to 32 breakpoints.
2. Navigate to Configure/Setup > Setup > Linearization and click on Characteristic to preview
the characteristic curve of the vessel breakpoints. (The conversion will function correctly
even if not all breakpoints are entered, but the curve will display correctly only if all
breakpoints are entered.)
3. In Step 5 Summary, check parameter values. Click on the appropriate step menu to return
and revise values, or click on a different menu to exit Quick Start.
Changing parameter settings using AMS Device Manager
Note
For a complete list of parameters, see Parameter Reference (Page 107).
For more detailed explanations of the parameters listed below see the pages referenced.
1. Adjust parameter values in the parameter value field in Configure/Setup view, then click
on Apply to write the new values to the device. The parameter field will display in yellow until
the value has been written to the device.
2. Click on OK only if you wish to update all parameters and exit AMS.
Remote operation
7.2 Operating via AMS Device Manager
SITRANS LR250 with mA/HART
Operating Instructions, 10/2019, A5E32220602-AG 89