Allows you to enable a maintenance limit.
Values
* Timer off
On - no limits
On - limit 1
On - limits 1 and 2
On - limit 2
1. First set the limit values in Maintenance Required Limit/Maintenance Demanded Limit.
2. Select the desired Maintenance Alert Activation option
8.7.14 Total Service Interval (4.16.6.)
Note
The device always operates in years. Changing the units affects only the parameter view of the
Service Interval parameters in SIMATIC PDM.
User-configurable recommended time between product inspections
Values
Units
1)
: hours, days, years
Range: 0 to 20 years
Default: 1 year
1)
Units are selectable only via SIMATIC PDM.
8.7.15 Maintenance Status (MAINT. STAT) (4.16.7.)
Read only. Displays the status of the Maintenance limits.
Open the menu View – Display Status, click on the Maintenance tab and check the Service
Schedule Status window.
8.7.16 Acknowledge Status (ACK. STATUS) (4.16.8.)
Read only. Acknowledges the current limit status.
Open the menu View – Display Status, click on the Maintenance tab and check the Service
Schedule Status window.
8.7.17 Acknowledge (ACK.) (4.16.9.)
Acknowledges the current limit status.
Parameter reference
8.7 Service (4.)
SITRANS LR260 with mA/HART