8.7.24 Total Calibration Interval (TOTAL CALIB. INTRV.) (4.17.6.)
Note
The device always operates in years. Changing the units affects only the parameter view of the
Calibration Interval parameters in SIMATIC PDM.
User-configurable recommended time between product calibrations.
Values
Units
1)
: hours, days, years
Range: 0 to 20 years
Default: 1 year
1)
Units are selectable only via SIMATIC PDM.
8.7.25 Maintenance Status (MAINT. STAT.) (4.17.7.)
Read only. Displays the status of the Maintenance Limits.
In SIMATIC PDM, open the menu View – Device Status, click on the Maintenance tab and
check the Calibration Schedule Status window.
8.7.26 Acknowledge Status (ACK. STATUS) (4.17.8.)
Read only. Acknowledges the status of the Maintenance limits.
In SIMATIC PDM, open the menu View – Device Status, click on the Maintenance tab and
check the Calibration Schedule Status window.
8.7.27 Acknowledge (ACK.) (4.17.9.)
Acknowledges the current limit status.
To acknowledge an alert via SIMATIC PDM:
1. Open the menu View – Device Status and click on the Maintenance tab.
2. In the Calibration Schedule Status section, click on Acknowledge Warnings.
To acknowledge an alert via the handheld programmer:
1. Press RIGHT arrow twice to open parameter view and activate Edit Mode.
2. Press RIGHT arrow to acknowledge the alert.
Parameter reference
8.7 Service (4.)
SITRANS LR260 with mA/HART
110 Operating Instructions, 11/2019, A5E32337683-AD